2.4.4 Performance Measurement
In order to obtain precise, repeatable and meaningful measurement results, it is absolutely necessary to
subject the test samples to a defined procedure consisting of a defined heating and reconditioning step
before a characteristic curve can be measured. The following procedure must be followed:
1) Heat the sensors from room temperature to 125 °C
2) Store them at 125 °C ±5 °C for at least 6 hours.
3) Allow the sensor to cool down to room conditions
4) Moistening:
a. Store sensor at 23 °C and 90 %RH between 4 and 6 hours
b. Take 2 measurements at 23 °C and relative humidity between 15 %RH and 95 %RH
5) Perform your characteristic curve measurement