Data Sheet

19
www.epluse.comv1.3 / Modification rights reserved
HTE501
3 POR
0: no POR since status 1 clear
1: POR occurred
This bit is cleared upon the Clear Status Register 1 command
2 Reserved -
1 Reserved -
0 CRC 1: checksum of the latest write transfer failed
Table 12: Status Register 1
BIT
NAME DESCRIPTION
15 NEW_MEAS New measurement (T, RH) since last readout available
14 NEW_T_MEAS
New T value since last readout available
Cleared upon start of T/RH readout
13 NEW_RH_MEAS
New RH value since last readout available
Cleared upon start of T/RH readout
12 NEW_TD_MEAS
New TD value since last readout available
Cleared upon start of TD readout
11 Reserved -
10 PERIODIC_MODE
Status of cyclic measurement
0: only measurements on demand
1: periodic mode active
9 Reserved -
8 Reserved -
7 Reserved -
6 Reserved -
5 Reserved -
4 Reserved -
3 Reserved -
2 Reserved -
1 RH_VALID
0: RH measurement faulty
1: RH measurement OK
0 T_VALID
0: T measurement faulty
1: T measurement OK
Table 13: Status Register 2