641
701-021
Cat. No. Description Price
701-021 POLARstar Optima computer-controlled fluorescence polarisation reader POA
2x extended wavelength PMTs (240nm to 740nm range)
2 filter wheels with 8 excitation and 8 emission filter positions
Supplied with 4 excitation and 4 emission filters*
High energy xenon flash lamp. Windows® based control software
and Microsoft Excel™ worksheet macro included
701-024 Reagent injection POA
* Excitation filters: 320nm, 355nm, 390nm, 485nm, 515nm, 544nm, 584nm and time-resolved (310nm to 370nm)
Emission filters: 405nm, 460nm, 538nm, 555nm, 590nm, 612nm, and time-resolved, please specify when ordering
Specification
Standard fluorescence Spectral range (excitation and emission)
240nm to 740nm. Dynamic range 8
decades. Sensitivity 0.9fmol/well
fluorescein
Time-resolved fluorescence Spectral range (excitation and emission)
240nm to 740nm. Dynamic range 7
decades. Sensitivity 50amol/well
europium
Fluorescence polarisation Spectral range (excitation and emission)
350nm to 740nm. Dynamic range 5
decades. Sensitivity 500fmol/well
fluorescein
Absorbance Spectral range 340nm to 740nm. Dynamic
range 0.000 to 3.000 OD. Reproducability
0.010 OD for 0–2 OD range, 0.030 OD for
2–3 OD range
Luminescence 0.5 fmol/well ATP
Light source High-energy Xenon flash lamp
Detector Side window, current type photomultiplier
tube
Filters Filter wheels for 8 excitation and 8
emission filter positions
Gain control Software-selectable Gain 1:65535 in 4095
steps. Automatic Gain calculation
Shaking Linear and orbital shaking
Programmable shake time and width
Reagent dispensers Two built-in syringe pumps access
measurement position. Variable injection
speed. Individual injection volumes for
each well. Reagent backflushing
Incubation Incubation range from ambient +5ºC to
45ºC in 0.1ºC steps
Temperature stability, ºC ±0.2
Temperature gradient, ºC <0.5
Microplate formats Up to 384, Standard and TR fluorescence
up to 1536
Reading time Dependent on test definition. Shortest
reading time 14 seconds for 96 well plate,
28 seconds for 384 well plate
Computer interface RS232, 9600baud, binary communication
protocol
Power requirements 100/115/230V, 50/60Hz, maximum
200VA
Dimensions (wxdxh), mm 440x480x270
Weight, kg 28
BMG, POLARstar Optima,
Microplate Fluorescence
Polarisation Reader
The POLARstar Optima is a fully automated
microplate reader that incorporates five different
measurement principals: standard fluorescence,
time-resolved fluorescence, fluorescence polari-
sation, absorbance (OD) and high performance
luminescence (2,000,000 cps). The unit can be
configured with on-board reagent dispensers
and an incubation chamber. The wavelength
range for standard and TR fluorescence is 240nm
to 740nm (extended wavelength to 900nm is
available, please enquire for details).
■
Simultaneous dual emission detection system
■
Instant top and bottom optics switching
■
Built-in syringe pumps access measurement
position
■
Programmable multiple fluorophore analysis
■
Precise temperature control and flexible
shaking
■
Reads up to 384 well plates (1536 standard
fluorescence and TR fluorescence)
■
Sensitivity 500fmol/well fluorescein
(polarisation mode), 0.9fmol/well fluorescein
(intensity mode)