Specifications

Order by phone 01342 826836 Order on-line www.ecomcat.co.uk
Order by fax 01342 826771 Website www.jencons.co.uk
Microplate Equipment
Readers
Take a look at our latest products on www.ecomcat.co.uk
629
Dynex Technologies, Spectra MR Scanning Microplate Reader,
continued…
The Spectra reader is supplied with Revelation software, cables and operating manual.
Specification
Light source Xenon flash lamp
Wavelength range 200–1,000nm Tunable Monochromator
Wavelength bandwidth 2nm ±0.3nm
Wavelength accuracy ±1nm
Wavelength resolution 1nm
Photometric linearity ±2% from 0.0 A to 3.0 A
Photometric range 0.0 A to 4.0 A
Photometric resolution ±0.001 A
Photometric accuracy ±0.005 A or <1.0% from 0.0 to 2.0 A
2.0% from 2.0 A to 3.0 A
Photometric precision (96 well plate) <0.005 A SD or <1% CV (whichever is greater)
between 0.0 and 2.0 A; <2.0% CV between 2.0 and 3.0 A
Stray light < 0.03% @ 230nm
Read time 96 wells < 20 seconds, typical (single wavelength)
384 wells <60 seconds, typical (single wavelength)
Scanning speed >90nm /min
Temperature range Ambient +3°C to 45°C
Temperature control accuracy ±1.0°C at 37°C (average of 96 wells)
±0.5°C well-well variation at 37°C
Shaking Linear, 3 user selectable modes
Plate types 6, 12, 24, 48, 96, 384 well
Reading methods Endpoint, kinetic, scanning
Interface USB
Computer requirements USB port, Windows® 2000 or XP with 500MHz
64MB RAM, 4 GB hard drive
Dimensions (dxwxh), mm 443x401x224
Electrical requirements 100-240V
Weight, kg 17.7
Cat. No. Description Price
448-025 Spectra MR microplate reader supplied with Revelation software, cables POA
and operating manual
Accessories
Cat. No. Description Price
448-023 Validation procedures (IQ,OQ,PQ) POA
For other options and accessories, please enquire on 01342 826836.
Spectra Verification Plate
The Spectra MR verification plate features:
Verification process is fully automated within Revelation software options to eliminate human error
The process tests if the following meet specifications; Linearity (absorbance accuracy), Precision
(reproducibility) and Wavelength accuracy in both visible and UV ranges
Spectral response is verified to ensure the complete spectral range is being detected as well as
stray light rejection
NIST traceability is provided for each filter in the verification plate along with a protective
storage case
Re-calibration service for annual recalibration
Cat. No. Description Price
448-024 Spectra MR verification plate POA
For information on UV microplates, please contact our Technical Office on 01342 826836.
448-024
New