Product specifications

Smart LT Manual Dynatronix, Inc.
198-0853-02 Rev F Page 95 of 126


.

.

.

.

Enter the local calibration menu from the calibrate configuration menu. This
menu is used to calibrate the current and voltage meters and the power supply
output.



Return to Calibrate
Configuration Menu
Discard any changes made
during this calibrate
session (only shown if
changes were made).
Save any changes made
during this calibrate
session (only shown if
changes were made).
Local Calibration
Menu (Pg 2 of 2)



The upper voltage output is calibrated at 95% of the
voltage rating. In operate mode the output should be
adjusted to 95% of the voltage rating by adjusting the
calibration value shown on this screen. Refer to the
calibration procedure for details on calibrating.


In operate mode the upper voltage output calibration
value is used to output 95% of the voltage rating and
the meter reading calibration value shown is
automatically adjusted to achieve the proper meter
reading. Refer to the calibration procedure for
details on calibrating.



The lower voltage output is calibrated at 10% of the
voltage rating. In operate mode the output should be
adjusted to 10% of the voltage rating by adjusting the
calibration value shown on this screen. Refer to the
calibration procedure for details on calibrating.


In operate mode the lower voltage output calibration
value is used to output 10% of the voltage rating and
the meter reading calibration value shown is
automatically adjusted to achieve the proper meter
reading. Refer to the calibration procedure for
details on calibrating.