Product specifications

Page 94 of 126 198-0853-02 Rev F
A.6.5.1 Local Calibration Menu




Enter the local calibration menu from the calibrate configuration menu. This
menu is used to calibrate the current and voltage meters and the power supply
output.

Local Calibration
Menu Page 2
Local Calibration
Menu (Pg 1 of 2)


.


Select YES to reset all
local calibration
settings to default
values.
The upper current output is calibrated at 95% of the
current rating. In operate mode the output should be
adjusted to 95% of the current rating by adjusting the
calibration value shown on this screen. Refer to the
calibration procedure for details on calibrating.

.


In operate mode the upper current output calibration
value is used to output 95% of the current rating and
the meter reading calibration value shown is
automatically adjusted to achieve the proper meter
reading. Refer to the calibration procedure for
details on calibrating.


.


The lower current output is calibrated at 10% of the
current rating. In operate mode the output should be
adjusted to 10% of the current rating by adjusting the
calibration value shown on this screen. Refer to the
calibration procedure for details on calibrating.

.


In operate mode the lower current output calibration
value is used to output 10% of the current rating and
the meter reading calibration value shown is
automatically adjusted to achieve the proper meter
reading. Refer to the calibration procedure for
details on calibrating.