Product specifications

Page 36 of 126 198-0853-02 Rev F
Digital Input Start Configuration:
“d.in Strt” = “OFF”
“d.in Strt” = “on”
No action is taken when the inhibit input changes
from a hardware disable state (INHIBIT lamp is on)
to a hardware enable state (INHIBIT lamp is off).
Use this setting when the inhibit input is used as a
safety interlock.
When the inhibit input changes from a hardware
disable state (INHIBIT lamp is on) to a hardware
enable state (INHIBIT lamp is off) the operate state
changes from standby to operate (a new cycle is
started if a cycle was not paused, or resumed if the
cycle was previously paused). Use this setting
when the inhibit input is used for remote
operate/standby control.
CAUTION: This setting will cause the unit to go to
operate mode any time the INHIBIT lamp goes from
an on to a off state.
Digital Input Stop Configuration:
“d.in StoP” = “PAUS”
“d.in StoP” = “End”
When the inhibit input changes from a hardware
enable state (INHIBIT lamp is off) to a hardware
disable state (INHIBIT lamp is on) the currently
running cycle is paused and the output is disabled.
If “d.in Strt” = “on” the cycle will resume when the
inhibit input changes to a hardware enable state.
Use this setting to pause a currently running
timed cycles when the inhibit input disables the
output.
When the inhibit input changes from a hardware
enable state (INHIBIT lamp is off) to a hardware
disable state (INHIBIT lamp is on) the currently
running cycle is terminated and the operate state
changes to standby. If “d.in Strt” = “on” a new
cycle will start when the inhibit input changes to a
hardware enable state. Use this setting to end a
currently running timed cycle when the inhibit
input disables the output.
The digital input configuration can be changed in the “Digital Input System Menu”. Use the
following procedure to configure the digital input. The “System Configuration Menu” is located
within the “Unit Configuration Menu (see 6.3.Unit Configuration). See A.6.4.System Configuration
Menu and A.6.4.2.Digital Input System Menu for more information regarding menu navigation.
1) From the “Unit Configuration Menu” rotate the current encoder until the display shows
“CFg” “SyS”, and then press the voltage encoder switch to enter the “System
Configuration Menu”.
2) From the “System Configuration Menu” rotate the current encoder until the display
shows “SyS” “d.in”, and then press the voltage encoder switch to enter the “Digital Input
System Menu”.
3) From the “Digital Input System Menu” rotate the current encoder until the display shows
“d.in” “Strt”, and then press the voltage encoder switch to view and change the digital
input start configuration.
4) The digital input start configuration is indicated by displaying d.in CyCL Strt” on the top
display and the digital input start configuration in the bottom display. The SET lamp is
blinking on and off to indicate that a setting is displayed. Rotate the current encoder to
change the digital input start configuration. Select “OFFif the inhibit input will be used