Product specifications

Page 20 of 126 198-0853-02 Rev F
Units using Only Voltage Control and Readback should be wired as shown below. The current
level is set to maximum using a resistor.
Typical Analog Control Wiring (Voltage Control Only)
Connector Signals
1 shield gnd
2 no connection
3 dc common
4 +12 (50ma)
5 -volt cntl in
6 +volt cntl in
7 -curr cntl in
8 +curr cntl in
9 +volt rb out
10 -volt rb out
11 + curr rb out
12 - curr rb out
1
12
6
Analog Control Connector
IN
OUT
Controller (Voltage)
4-20ma, 0-5V or 0-10V
+
-
+
-
Earth
Gnd
Shields
Shields
Approximately
560 Ohm
Resistor, ½ Watt
Jumper
Alternative to the resistor:
The current can be adjusted to
the desired value in Local
control. Then the current setting
source can be changed to host
using the procedure in section
6.3.3. Finally when the unit is
changed to Remote control with
the RMT switch, the current
setting will be retained as long
as there is not a host to change
it.