Product specifications
Diamond 36kW Manual Dynatronix, Inc.
198-1600-00 Rev F Page 73 of 146
3. From the module read back calibration menu rotate the current encoder until the display
shows “CAL” “End”. Press the voltage encoder switch to exit the module read back
calibration menu and return to the “Calibrate Configuration Menu”.
If changes were made to any calibration values you will be prompted to store the changes. Select
“yES” to accept the changes or “no” to cancel the changes to the calibration values
10.2.4. Analog Calibration (Optional)
If the optional analog control interface is installed in the power supply, follow this procedure to
calibrate it.
NOTE: The existing calibration values can be viewed without changing them by leaving the supply
in LOCAL mode (RMT/LCL lamp is off) and by selecting “no” if prompted to store changes.
NOTE: When calibrating the analog interface, there will be no current or voltage on the output of
the power supply. This procedure calibrates the analog interface circuitry.
NOTE: If an option card is not installed (as determined by the unit configuration, see
6.3.8:Hardware Options Configuration) the option card calibration menu will display “no” “OPt”.
The calibration steps will not be available.
10.2.4.1. Analog Calibration Connections
10.2.4.2. Analog Calibration Procedure
This sequence sets the calibration values for the optional analog interface.
.
1. From the “Calibrate Configuration Menu” rotate the current encoder until the display
shows “CAL” “OPt”. Press the voltage encoder switch to enter the “Option Card
Calibration Menu” (see A.7.5: Calibrate Configuration Menu and A.7.5.2: Option Card
Calibration Menu for more information regarding menu navigation). The calibration
menu contains a “dFLt” sub menu to load default calibration values (see
Connector Signals
1 shield gnd
2 no connection
3 dc common
4 +12 (50ma)
5 -volt cntl in
6 +volt cntl in
7 -curr cntl in
8 +curr cntl in
9 +volt rb out
10 -volt rb out
11 + curr rb out
12 - curr rb out
1
12
+
-
+
-
IN
OUT
Calibrator
4-20ma, 0-5V or 0-10V
6
Connections for
current calibration
Connections for
voltage calibration
Analog Control Connector
Connections for Analog Calibration