Product specifications

Diamond 36kW Manual Dynatronix, Inc.
198-1600-00 Rev F Page 37 of 146
2) From the “System Configuration Menu” rotate the current encoder until the display
shows “SyS” “Strt”, and then press the voltage encoder switch to enter the “Unit Start
System Menu”.
3) From the “Unit Start System Menu” rotate the current encoder until the display shows
“Strt” “CtrL” (“Start Control”), and then press the voltage encoder switch to view and
change the power on configuration.
4) The power on configuration is indicated by displaying Unit Strt CtrL” on the top display
and the selected power on configuration in the bottom display. The SET lamp is
blinking on and off to indicate that a setting is displayed. Rotate the current encoder to
change the selected power on configuration. Select “Stby” to configure the unit to
always power on in standby mode. Select “LASt” to configure the unit to power on in
the same mode it was in when it was powered off. Once the desired configuration is
selected press the voltage encoder switch to accept the value and return to the “Unit
Start System Menu”.
5) From the “Unit Start System Menu” rotate the current encoder until the display shows
“Strt” “End”. Then press the voltage encoder switch to exit the “Unit Start System
Menu” and return to the “System Configuration Menu.
6) From the “System Configuration Menu” rotate the current encoder until the display
shows “SyS” “End”. Then press the voltage encoder switch to exit the “System
Configuration Menu” and return to the “Unit Configuration Menu.
7) If any settings were changed in the “System Configuration Menu” you will be prompted
to store the changes. Select “yES” to accept the changes or “no” to cancel the
changes.
6.3.5. Power Fail Alarm Configuration
The power supply can be configured to report error codes if power is interrupted while the supply
is in operate mode. If the power failure alarms are enabled the appropriate error code (error
codes 94 or 95) is set when power fails during operate mode. The error code will remain active
until it is manually cleared, even if the unit is powered on and off multiple times. See Appendix
C:Error Codes for a complete list of error codes. See 6.5.Error Messages for details regarding
error message display and manual clearing of error codes.
Power failure alarms can be enabled or disabled in the “Unit Start System Menu”. Use the
following procedure to enable or disable power failure alarms. The “System Configuration Menu”
is located within the “Unit Configuration Menu (see 6.3.Unit Configuration). See A.7.4: System
Configuration Menu and A.7.4.1: Unit Start System Menu for more information regarding menu
navigation.
1) From the “Unit Configuration Menu” rotate the current encoder until the display shows
“CFg” “SyS”, and then press the voltage encoder switch to enter the “System
Configuration Menu”.
2) From the “System Configuration Menu” rotate the current encoder until the display
shows “SyS” “Strt”, and then press the voltage encoder switch to enter the “Unit Start
System Menu”.