Product specifications
Diamond 36kW Manual Dynatronix, Inc.
198-1600-00 Rev F Page 133 of 146
analog input. The “s” command can be used by the host to change other settings when the control
type is set to “3”.
4. If the power up mode is set to 1 and the unit was powered down in operate mode, then the unit will
display a count down when it powers up (also reported by the “d” readings command). If no action
is taken the unit switches to the operate state when the countdown expires. If the operate/standby
switch is pressed at any point prior to the countdown expiration the countdown will terminate and
the unit will remain in standby mode. If the host requests (using the state command) operate or
standby prior to the countdown expiration the countdown is terminated and the requested mode is
set.
5. Each time the unit configuration password is changed with this command the value is written to
non-volatile memory. Periodic writes to this value could result in premature failure of the associated
memory device. Leave this field blank unless the code needs to be changed. All other parameters
in this command that are stored to non-volatile memory are written to non-volatile memory at power
down and can be safely changed at periodic intervals.
6. When the unit configuration password is changed the slave may take up to 100ms to acknowlege
the request (write to non-volatile memory).