Product specifications

Diamond 36kW Manual Dynatronix, Inc.
198-1600-00 Rev F Page 105 of 146
.
.
.
----------------------- Standby Mode : Volt Setting------------------
----------------------- Operate Mode : Volt Gain---------------------
.
.
.
.

.




Return to Calibrate
Configuration Menu
Discard any changes made
during this calibrate
session (only shown if
changes were made).
Save any changes made
during this calibrate
session (only shown if
changes were made).
Module
Calibration Menu
(Pg 2 of 2)



The CL04 (module A) and CL08 (module B) menus
show the voltage reading gain value. When this
menu is viewed in standby mode (as shown to the
left) the voltage setting in volts that will be used for
calibration is shown. The output voltage can be
adjusted to an appropriate level for calibration.


The CL04 (module A) and CL08 (module B) menus
show the voltage reading gain value. When this
menu is viewed in operate mode (as shown to the
left) the gain value is shown. The gain value is
adjusted to increase or decrease the module output to
match the output voltage set in CL04/CL08 standby
mode.
Enter the Module calibration menu from the calibrate configuration menu.
This menu is used to calibrate the current and voltage meters and the module
output.
CAUTION: Each 18kW output module must be calibrated individually.
There are two module calibration menus, one for module A and one for
module B.
.



The FLtA (module A) and FLtB (module B) menus
show the reading filter time in seconds. In most
cases the shortest filter time will work best. Longer
filter times provide more filtering, but the meters will
not react as quickly to changes in output.