Product Manual

Dynatron 900
+™
12
Intermittent Cycle
Longer Ramp Time
STEPLESS PROGRESSIVE
1. Prepare the patient for the traction treatment. Make sure all belts and harnesses are securely
attached.
2. Plug in the remote stop cable and hand it to the patient.
3. Turn on the Dynatron 900+ device.
4. Set Treatment Time: 1-99 min.
5. Set Hold Force Timer: 1-99 sec.
11. Set Rest Force Timer: 1-99 sec.
12. Set Hold Force: 4 200 lbs. or 2 - 91 kg.
13. Set Rest Force: 0 197 lbs. or 0 - 90 kg. Must be set
lower than Hold Force value.
14. Adjust Speed Slow (minimum setting)
15. Press START
NOTE: AS A SAFETY PRECAUTION, IT IS NECESSARY TO RESET BOTH THE HOLD AND
REST FORCES WHENEVER THE POWER SWITCH IS ARE TURNED ON. FAILURE TO DO
SO WILL RENDER THE DEVICE INOPERABLE.
SPECIAL USAGE PROCEDURES
ADJUST HOLD/REST FORCES AFTER A TREATMENT IS IN PROGRESS
1. Once the initial Hold Force is reached during a treatment, set the Hold or Rest Force parameters
to the new values.
2. Press the Start Key. The change will not take place until the next full Hold/Rest cycle.
3. In static mode, the Dynatron 900+ device will not accept any change until the Reset Key is
pressed.