MAINTENANCE AND TROUBLESHOOTING
SMARTRAC | OPERATOR’S MANUAL REV. 0 | 4/06/2018 26
Patient Stop Switch
e Patient Stop must be held by the patient during a treatment. e patient can press the switch to stop the treatment
immediately for any reason.
e Patient Stop switch must be properly connected to the console. Treatment cannot start unless the Patient Stop Switch
is connected.
e switch should be handed to the patient and its use explained before treatment is started. Make sure the stop switch can
be used by the patient during the whole treatment. If it is pressed during treatment, the traction force will be reduced to the
minimum of 3.5 lbs (15 Newtons) at the maximum change over speed. At the same time a buzzer will sound.
reshold Cervical Traction
For Cervical Traction, forces greater than 45 lbs. (200 Newton) are not suitable. When setting the traction force, a warning
will be issued when exceeding this limit. To ensure safe use, a force higher than 45 lbs. (200 Newton) can be selected only in
case it is conrmed by the professional user that this force will not be applied with cervical traction.
Technical Maintenance
Electrical safety of the device relies on a properly grounded electrical connection via the power cord. It is therefore necessary
to have this connection checked annually. To ensure continued compliance standard, this unit should be adjusted and safety
tested once each year. is may be carried out by your supplier, or by another agency, authorized by the manufacturer. If a
problem occurs or technical service is requested please contact Dynatronics (800) 874-6251. It is also recommended that a
service history record is maintained. In some countries this is may be obligatory.
CAUTION
Electrical safety of the device relies on a properly grounded electrical connection via the power cord. It is therefore
necessary to have this connection checked annually.
Use of controls or adjustments or performance of procedures other than those specied herein may result in
hazardous exposure to ultrasonic energy.
WARNING
is unit operates with high voltages. No attempt should be made to disassemble the unit. Maintenance and repair
should be carried out by authorized personnel only. Dynatronics will not be held responsible for the results of
maintenance or repairs by unauthorized persons.