Product Manual

HIGH VOLT INSTRUCTIONS
DYNATRON® 25 SERIES™ | OPERATOR’S MANUAL 42
1. Press the HI VOLT key.
e High Volt channel and the default settings for High Voltage electrode pulsed stimulation are automatically
selected. Using the dedicated High Volt (HV) channel attach leads and place electrodes on the patient now. .
Default Setting
Treatment ........ High Volt Pads Treatment
Duty Cycle .............................. Continuous
Time Electrodes .......................10 minutes
Polarity ..........................................Negative
Pulse Rate................................ High Range
Range: ...............................High 80-120 Hz
If you wish to use the default settings, increase the intensity to the desired level, and press START. To customize the
settings, follow steps 2-8.
ELECTRODE PADS TREATMENTS are timed, and a treatment time in MINUTES is entered at the start of the
treatment. e timer counts DOWN for electrode treatments.
2. Choose the POLARITY
Press the Polarity toggle key to choose Polarity. Polarity options are Negative, Positive, and Bipolar.
3. Change the treatment TIME, if desired.
Use the TIME UP/DOWN ARROW keys to change the TIME for an Electrode treatment. Using the MODE toggle
key, be sure TIME is illuminated in the MODE window when entering TIME.
4. PULSE RATE RANGE.
Select a PULSE RATE RANGE (High 80-120 Hz or Low 1-10 Hz) or set a single pulse rate by pressing the MODE
toggle key and selecting RATES from 1-120 Hz. Use the arrow keys next to RATE-START and RATE-STOP to set
the Pulse Rate Range. ese ranges may be changed for a single treatment if desired, or press and hold the START
key to set new default settings to be applied to all future treatment setups. e PULSE RATE RANGE must be set
before pressing START.
5. CONTRACTION (ON) and REST (OFF) TIMES
FUNCTION KEY: To access CONTRACTION and REST settings, press the FUNCTION key.
Press the CONT/REST toggle key one or more times to select contraction/rest (Duty) cycle times. Available options
include 10/10, 10/30, 10/50, CONT (Continuous). and Custom. e setting of 10/30, for example, means a 10-second
contraction time followed by a 30-second rest time.