7. SAVE DEFAULTS
If the treatment you have just set up is a frequently used, you can save the treatment parameters as new defaults by
pressing and holding the START key until a beep sounds indicating the treatment parameters have been saved. e
next time you select the modality, these parameters will be selected automatically.
8. MODIFY SETTINGS
Treatment settings can be modied while the treatment is in progress except for a CONSECUTIVE HIGH/LOW
treatment. If the TIME setting on a CONSECUTIVE HIGH/LOW treatment is altered, the treatment will be aborted
and default to an ALTERNATING HIGH/LOW treatment.
• FREQUENCY RANGE. Use the Range Toggle key to select a dierent frequency option (High, Low, Alternating
High/Low, Consecutive High/Low, or Fixed). Fixed Frequency and Consecutive High/Low cannot be selected
aer treatment is started.
• TARGET/SWEEP/STATIC. Use the Target Toggle key to select Target, Sweep, or Static (for IFC only).
• TIME. Use the Time Arrow keys to increase or decrease the treatment time.
• INTENSITY. Use the Intensity Arrow keys to increase or decrease the intensity.
• TARGET. Relocate the treatment site by touching the TARGET PAD at any time during the treatment when
TARGET has been selected.
9. STOP.
When the treatment time has elapsed, the current to the patient stops and a tone sounds signaling the end of a
treatment. Treatments in progress may be stopped at any time using one of the following methods.
Stop One Treatment Only. Press and hold the FUNCTION key and press STOP. is stops only the treatment in focus.
Stop All. Press the STOP key. All treatments at all channels will stop.
Stop Time. Reduce the treatment time using the Time arrow key. e output at the selected channel is stopped,
and the device then displays the parameters for the next treatment that remains in progress (if there is an active
treatment).