INTERFERENTIAL / PREMODULATED INSTRUCTIONS
DYNATRON® 25 SERIES™ | OPERATOR’S MANUAL 26
If you wish to change or customize the treatment settings, proceed through the following steps:
2. Customize TIME
e default treatment time is displayed at 10:00 min. Use the TIME arrow keys to increase or decrease the treatment time.
3. Customize FREQUENCY (optional)
Default HIGH and LOW Frequency Settings
• HIGH range is 80 to 150 Hz.
• LOW range is 0 to 10 Hz.
• ALTERNATING range alternates every 30 seconds between HIGH and LOW, beginning with LOW.
• CONSECUTIVE HIGH/LOW. During the rst half of the treatment time the High frequency range is delivered.
During the second half of the treatment time the Low frequency range is delivered.
• FIXED. A FIXED treatment automatically defaults to a STATIC setting. Press the toggle key under MODE until
FREQ is illuminated. A FIXED treatment can be set between 4,000 and 10,000 KHz. Selections progress from
4,000 KHz to 10,000 KHz in 1,000 KHz increments. Once the FIXED rate is set, return to the TIME display by
using the MODE toggle key.
e default High and Low frequency settings may be changed for a single treatment if desired, or new default settings
may be saved to apply to all future treatment setups.
• Press the toggle key under the MODE display. Select FREQ (Frequency).
• Press the toggle key under the RANGE display. Select a HIGH or LOW frequency range.
• e HIGH Frequency will appear on the le-hand side of the Treatment Display Screen. e LOW Frequency
will appear on the right-hand side of the Treatment Display Screen. Changes to the upper and lower limits are
NOTE: ALTERNATING and HIGH/LOW options may be selected aer pressing START. However, Consecutive
and Fixed options must be selected before pressing START.
NOTE: Make any desired changes to the treatment time before selecting Consecutive. Treatment time changes
made aer selecting Consecutive will cause the treatment to revert to an ALTERNATING HIGH/LOW treatment.
NOTE: If INTENSITY has been set during the setup of a FIXED treatment option and a change is made to any
other treatment option, INTENSITY will automatically return to “0.”