Product Manual
GENERAL SPECIFICATIONS
DYNATRON
®
125 | OPERATOR’S MANUAL REV. 3 | 8/29/2017 36
Ultrasound Head
• Ultrasound heads should be cleaned with warm water. Always keep the head free from gel buildup. Alcohol may be
used to sterilize the soundhead.
• Do not use ice water for cooling soundheads.
• Do not allow soundheads to overheat repeatedly. is could result in thermal shock to the crystal. Damage of this
type is not covered by the warranty.
• Do not drop the unit or the soundheads as severe damage will occur.
Suggested Maintenance Schedule
Service To Be Performed Annually By A Technician:
• Annual Ultrasound calibration should be performed by a qualied technician.
• Inspect soundhead connectors on unit and on soundhead.
Maintenance Performed By User:
1. Inspect accessories daily for wear and damage. Examine cables and connectors on the cables for any visible sign of
wear or damage. Replace accessories as needed.
2. Examine Ultrasound heads periodically for cracks which may allow ingress of conductive uid.
3. If a machine or soundhead is dropped, or if it sustains damage due to lightning, severe power surge, or other incident
that could cause damage to electronic components, the device must be examined by a Dynatronics technician before
being returned to clinical use.
4. Even if the machine is functioning properly, you can send it to Dynatronics for preventative maintenance service for
a nominal charge; call for pricing.
5. Inspect device air vents periodically to ensure air ow is not blocked. An ordinary household vacuum hose may be
used to clean dust from air vents externally.
6. ere are no serviceable parts in the Dynatron 125. DO NOT attempt to unscrew or tighten the Ultrasound probe
heads. Immediately report any device malfunction to Dynatronics Customer Service Department.
WARNING
Hazardous electrical output. To reduce the risk of electrical shock, do not remove cover. Refer servicing to
qualied service personnel.