BATTERY OPERATION
DYNATRON
®
125 | OPERATOR’S MANUAL REV. 3 | 8/29/2017 33
treatment has ended. e yellow BATTERY ICON will appear in the lower le-hand corner of the Operations
Screen, indicating the battery level.
5. When the available battery power becomes too low to continue operating the device, the following message will
appear: “ERROR: BATTERY LEVEL TOO LOW FOR TREATMENT OPERATION. CHARGE IMMEDIATELY!”
e treatment intensity will ramp down, any treatments that were running at the time will stop, and the device will
shut down. Before battery operation can continue, the battery must be recharged.
Battery Requirements
12 volt and at least 5 amps hours.
Battery adaptor cord must match the plug end of the battery pack. e barrel plug end must match the 0.325” barrel
jack adaptor plug on the Dynatron 125.
e cord needs to be a minimum of 18 AWG gauge wire. 14-16 AWG gauge wire will work as well with a 5 amp fast
blow fuse.
Battery Life
e length of time that a unit can be used with a battery pack is dependent on several factors:
e amperage of the battery pack. Larger amperage will provide longer use.
e amount of charge remaining on the battery.
As a general rule, the unit may be run continuously for 30 minutes to several hours depending on these factors.
CAUTION
Follow battery manufacturer’s instructions for usage and care. When disposing of a used battery, comply with the
laws and procedures required in your area.
Note: If using a smaller gauge wire (20 AWG and up) the BATT LOW error is possible when the battery is not low.