BATTERY OPERATION
DYNATRON
®
125 | OPERATOR’S MANUAL REV. 3 | 8/29/2017 32
Battery
Operation
All Dynatron 125 devices are manufactured with battery capabilities allowing deliver of treatments wherever power may be
unavailable or unreliable. To use the optional battery, do the following:
1. It is recommended that a battery be charged for 24 hours prior to
operating the Dynatron 125 device. DISCONNECT the battery
charging cable from the battery while it is in use for treatment.
2. Plug the battery adapter into the jack labeled “BAT-INPUT 12V-DC”
on the back of the Dynatron 125 console. e green BATTERY ICON
will appear in the lower le-hand corner of the Operations Screen.
3. Set up and deliver treatments.
4. When battery power is reduced to approximately 11 volts, a low battery warning will be displayed in the lower
right-hand corner of the Treatment Display Screen “CAUTION: BATTERY LEVEL GETTING LOW!” e treatment
can continue; however, there will not be enough power to set up and deliver another treatment when the current
Use ONLY a Battery Meeting Specications
Before purchasing or using an existing battery with the Dynatron 125 device, refer to the section titled “Battery
Requirements” in this manual.
Only use a battery that CANNOT be recharged while it is in use. Disconnect the battery charger from the
AC power source before using the battery to supply power to this device.
Battery Available
Caution, Low Battery
Battery Level Too Low.
Charge Immediately!