ULTRASOUND PROBLEM SOLVING
DYNATRON
®
125 | OPERATOR’S MANUAL REV. 3 | 8/29/2017 21
Ultrasound
Problem Solving
Whirlpool Treatments
If you are treating in a whirlpool, you may nd that the temperature in the whirlpool approaching 103˚F, causing the
overheated soundhead caution to appear on the Operations Screen. is is a cautionary warning only to let you know that
you are approaching the temperature limit. You may, however, continue with the treatment at this level. If your whirlpool
temperature is hot enough to cause the treatment to stop, you will need to adjust the temperature of the whirlpool.
Soundhead Temperature Too Cold
If the soundhead has been sitting in a very cold room or vehicle, it could be too cold to operate when plugged into the console.
e keypad may not respond to key presses and the device will not function until the soundhead is suciently warmed. You
must raise the temperature of the soundhead to about 60 degrees F in order for the machine to recognize that the soundhead
is present and to proceed with setting up a treatment. You can accomplish this with any of the following methods:
1. Press the at face of the soundhead against the palm of your hand for 30 to 60 seconds to warm it slightly. is
usually provides adequate warmth to the crystal to raise the temperature to the minimum acceptable level. Once the
crystal reaches this level, you can proceed with treatment.
2. You can also place the soundhead in room temperature water to warm the crystal. However, do not place the
soundhead in very hot water when the crystal is this cold as it could damage the crystal.
No Soundhead
If the device cannot detect a soundhead during setup or during an ultrasound treatment, the error message “SOUNDHEAD
DISCONNECTED! CORRECT PROBLEM WITH SOUNDHEAD TO RESUME!” will appear on the Operations Screen.