ULTRASOUND INSTRUCTIONS
DYNATRON
®
125 | OPERATOR’S MANUAL REV. 3 | 8/29/2017 14
If you wish to use the default settings, increase the Intensity to desired treatment level and press START. If you wish to
customize settings, follow steps outlined below.
2. Choose the FREQUENCY.
Press the FREQUENCY display toggle key on the Operations Screen to select 1, 2, or 3 MHz. Any one of the three
Frequencies may be used with the 2 cm
2
, 5 cm
2
or 10 cm
2
soundheads.
3. Select the DUTY CYCLE
Press the DUTY CYCLE toggle key located on the Operations Screen to select one of the four available options: 10%, 20%,
50%, or 100% (Continuous) duty cycles.
4. MODIFY TREATMENT PARAMETERS
e following parameters can be modied at the time of set-up or while the treatment is in progress. : FREQUENCY,
DUTY CYCLE, TIME, INTENSITY, HEAD WARMING, and POWER DISPLAY of WATTS or W/cm
2
.
• HEAD WARMING
Access the Settings Screen by pressing the Settings Key in the lower right hand corner of the Operations Screen.
On the Settings Screen, press the WARMING key to toggle between Head Warming ON and OFF. When the
image turns “BLUE,” Head Warming is ON. When the display returns to black and white, Head Warming is OFF.
• POWER DISPLAY
Press the Setting Key on the Operations Screen. From the Settings Screen, select the appropriate toggle key to
change the POWER DISPLAY from WATTS to W/cm
2
.
5. TIME.
Touch the TIME display on the Operations Screen to access the TIME SCREEN. e default time is set for a
5:00 min. treatment. Time can be changed by using the TIME Up/Down arrow keys located to the right of the
TIME display. Aer selecting the treatment time, press the Return Key located in the lower right-hand corner
to return to the Operation’s Screen.
6. SAVE SELECTED PARAMETERS AS THE DEFAULT TREATMENT PARAMETERS.
To save the treatment TIME that has been selected for a current treatment as the Default TIME, touch the TIME
window on the Operations Screen. Press the image in the lower le-hand corner. e selected treatment TIME
is now saved as the Default for future treatments. Any changes made to Language, Head Warming, and Intensity
Display Units are automatically saved as Defaults.
7. RAISE THE INTENSITY.
On the Operations Screen, use the INTENSITY Up/Down arrow keys located on either side of the Power Display to
increase the power to the desired treatment setting. For patient safety and comfort, it is recommended that treatment
begins with .1 W/cm
2
, before increasing power to the desired level aer the treatment begins. Valid ranges are from
0.1 to 2.0 W/cm
2
(exceptions: valid ranges when using a 10 cm
2
head are from 0.1 to 1.0 W/cm
2
).