Product Manual

ULTRASOUND INSTRUCTIONS
DYNATRON
®
125 | OPERATOR’S MANUAL REV. 3 | 8/29/2017 13
Detailed Ultrasound Setup
1. Turn the main power switch at the back of the unit to the ON position.
e following Default Parameters will automatically appear in the Operations Screen:
Frequency ........................................1 MHz
Duty Cycle .............................. Continuous
Display ..............................................W/cm
2
Intensity .................................... 0.0 W/cm
2
Time..............................................5:00 min.
Other Default Parameters that are displayed on the Settings Screen:
Warming .............................................. OFF
Language.........................................English
Ultrasound Quick Setup
1. Turn ON the main power switch at the back of the unit.
2. FREQUENCY. Using the FREQUENCY toggle key, select 1 MHz, 2 MHz, or 3 MHz.
3. DUTY CYCLE. Using the DUTY toggle key, select 10%, 20%, 50%, or 100% (Continuous).
4. TIME. Change the treatment TIME, if desired, by pressing the time display. Use the up and down arrow
keys to select the desired time. Press return aer the time has been modied.
5. INTENSITY. Use the Up/Down arrow keys within the INTENSITY button to raise or lower the INTENSITY
to the desired level.
6. Press START.
7. MODIFY a treatment in progress, if desired. While the treatment is in progress, the following parameters
can be modied: FREQUENCY, DUTY CYCLE, TIME, and INTENSITY.
8. PAU SE . e PAUSE key may be pressed at any time aer a treatment has been started. e output will stop
and the timer will PAUSE.
9. STOP. Press the STOP key to stop the treatment.