DYNATRON® 125™ PHYSICAL FEATURES
DYNATRON
®
125 | OPERATOR’S MANUAL REV. 3 | 8/29/2017 9
c. Battery. is jack may be used to supply power to the device using an optional battery pack. More information
about the optional battery operation is provided later in this manual.
Ultrasound Error Messages
If an error occurs during any active treatment, the Dynatron 125 will sound a tone/beep. An Error Message will appear on the
Operations Screen. Below are samples of the error messages that may occur.
ULTRASOUND ERROR MESSAGES CAUSE
ERR 200
SOUNDHEAD DISCONNECTED! TREATMENT PAUSED.
CORRECT PROBLEM WITH SOUNDHEAD TO RESUME
Soundhead disconnected
ERR 201
THERMISTOR ON SOUNDHEAD IS MALFUNCTIONING!
PLEASE REPLACE SOUNDHEAD.
Soundhead has a
broken thermistor
ERR 202 SOUNDHEAD CURRENT TOO HIGH! Soundhead over current
ERR 203 SOUNDHEAD CURRENT TOO LOW! Soundhead under current
ERR 204
SOUNDHEAD IS TOO HOT!
OUTPUT HAS BEEN DISABLED TO ALLOW COOLING.
Soundhead too hot
ERR 205
SOUNDHEAD WAS JUST DISCONNECTED,
DISABLING POWER OUTPUT!
Soundhead disconnected
during calibration
ERR 206
PATIENT COUPLING.
PLEASE REAPPLY SOUNDHEAD TO PATIENT. Inadequate patient coupling.
ERR 207
SOUNDHEAD CALIBRATION DATA IS CORRUPT.
PLEASE RE-INITIALIZE SOUNDHEAD DATA.
US Probe calibration
data is corrupted
ERR 208
NO HEAD INTENSITY.
TREATMENT WILL NOT START.
PLEASE INCREASE INTENSITY.
Intensity not set - or -
Head disconnected.
Bad soundhead cord.
ERR 209
SOUNDHEAD NOT DETECTED!
CORRECT PROBLEM WITH SOUNDHEAD TO RESUME.
Head disconnected.
Bad soundhead cord.
ERR 210
ENTERED CALIBRATION DATA RANGE ERROR.
CORRECT PROBLEM BY RE-ENTERING DATA.
Soundhead parameters manually
entered incorrectly.