Manual
To receive a customized proposal,
contact us today at:
Dynatek Labs
105 East 4th Street
Galena, MO 65656
800.325.8252
1.417.357.6155
www.dynateklabs.com
Dynatek’s CDT-5 Coating Durability Tester Specications
CDT-5
Description Specification
Number of test samples 5
Tube configuration Straight, curved or bifurcated
Tube length (straight) 88 to ≤ 250 mm
Tube outer diameter ≈1 mm to ≤ 10 mm
Typical distention 0.3% to 18% of tube ID
1
Testing rate 1.2 to 150 Hz
Testing fluid PBS or distilled water
Fluid temperature Ambient to ≤ 45º
Flow rate 100 mL/min standard/sample
Test parameters Stent deflection, tube deflection, pressure, temperature, frequency
Automatic test control method Tube distention (with optional laser micrometer)
Laser micrometer Solid-state laser diode (optional; 780 nm, 3 mw)
Pressure transducer range 0-10 psig
Number of particle counters 5
Particle size range 5-1,000 microns
2
Max. particle concentration 15,000 p/mL @ 10 microns
Particle bin range 2-6 bins
System fluid filter 0.2 micron
Particle capture filters 5 (optional)
Particle capture filter size 1 micron (optional, customizable)
Controller Universal Controller, featuring the NI™ Single-Board RIO platform
Software Invitroflow software package, with intuitive user
interface to monitor and control the CDT-5
Capacitance tank 0.47 gal
Module Dimensions Power Weight
CDT-5 (base unit) 14 ¾” W X 20”D X 55”H 5.0 A 95 lbs
1
Depending on sample compliance
2
Based on use of appropriate standards
CDT-5™, CDT-20™, Coating Durability Tester™ and Invitroflow™ are trademarks of Dynatek Labs, Inc.