Service manual

Overview
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 87
DNI Nevada CuffLink
Adult/Pediatric Mode
Clinical Dynamics Smart Arm
Adult/Pediatric Mode
Neonate Mode
FastBP StepBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
Expected
Systolic
Expected
Diastolic
200/150 210 152 201 138
120/80 116 77 124 80
80/50 87 49 90 50
FastBP StepBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
Expected
Systolic
Expected
Diastolic
200/150 192 152 194 136
120/80 124 79 122 80
80/50 85 42 80 49
NeoBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
80/50 77 48
60/30 57 32
35/15 31 13