9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 85
Appendix B
Overview
This appendix includes simulators and settings that may be used to assess the performance of the NIBP module.
Readings should be obtained within +/- 10 mmHg of the expected measurement result. The following simula-
tors are used in this chapter:
Bio-Tek Instruments BP Pump
Bio-Tek Instruments BP Pump2
DNI Nevada CuffLink
Clinical Dynamics Smart Arm
NIBP Simulator Performance
FastBP
A dual lumen hose is required for FastBP.When you use FastBP, the reading is obtained during cuff inflation.
StepBP
StepBP is compatible with both single and dual lumen hoses.When you use StepBP, the reading is obtained
during step deflation.