Service manual

CHAPTER 3DISASSEMBLY PROCEDURES
48 9650-0801-01 Rev. A
Kapton tape
3M copper adhesive tape, or equivalent
1/2” nut driver
Large diagonal cutters
Strong glue, such as Loctite 420 or equivalent
Loctite needle tip dispenser
Safety Precautions
WARNING! SHOCK HAZARD!
Caution TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY
BEFORE YOU CONDUCT DEFIBRILLATOR TESTS OR REPAIRS.
Only properly trained technicians should service the unit.
The unit can contain deadly voltages even if the unit is turned off.
Make sure to discharge the unit before working with it.
Make sure you take the necessary precautions when working with static sensitive units. For
example, you must wear a conductive wrist strap (which touches your skin) connected to a
grounding mat and to the earth ground. You must remove the wrist strap when you discharge
high voltage or when you are working on energized equipment.
Modules
The Propaq MD unit contains the following nine (9) modules or assemblies:
Printer
Front Enclosure
Display
NIBP
EtCO
2
(Oridion
®
)
Monitor Board
Capacitor Bank
Defibrillator Board
SpO
2
These modules can be purchased as replacement parts from the ZOLL Technical Support
Department. See Chapter 5, "Replacement Parts" for information on ordering replacement
parts.