Service manual
Propaq MD Error Messages
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 45
PS DEFIB ENERGY DOWN
SHORT
Paddle set energy down
select button is stuck.
Replace the paddles.
PS DEFIB RECORDER
SHORT
Paddle set recorder button is
stuck.
Replace the paddles.
DEFIB CAP LEAKAGE Capacitor bank cannot hold
a charge.
Replace the capbank
module.
DEFIB USING DEFAULT
CALIBRATION
Calibration data lost. Replace the defib board
module.
DEFIB FATAL CHARGE
ERROR
Charging error. Replace the defib board
module.
DEFIB CHARGE ERROR Over voltage was detected. If problem persists, replace
the defib board module.
DEFIB LOW BATTERY Battery is too low to charge
defibrillator.
Charge the battery. Check
battery contacts and clean
if necessary.
DEFIB CHARGE TIMEOUT
FAIL
The battery did not produce
enough current to charge
capacitor.
Charge the battery. If
voltages are okay (from
battery and aux), replace
the defib board module.
DEFIB DISCHARGE TEMP Device is in a hot
environment or the
discharge circuit is always
on.
Move device to cooler area.
If problem persists, replace
the defib board module.
DEFIB FIRE FAILURE High impedance or low
impedance (short).
Replace pads. If problem
persists, check internal
connections. If still persists,
replace the defib board
module.
DEFIB PHASE FAILURE Phase I or phase II
waveform dropout.
Replace the defib board
module.
DEFIB NO ENERGY
DELIVERED
High impedance. No defib
energy was delivered.
Replace pads. If problem
persists, check internal
connections. If still persists,
replace the defib board
module.
PACER DEVICE FAILURE Generation of pacer pulse
failed.
Replace the defib board
module.
PACER CURRENT FAULT Lack of pacer current. Replace the defib board
module.
Error Message Explanation Technical Action