Functional Verification Tests
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 33
IBP Test
Tools Needed IBP patient simulator (with accuracy better than +/- 1 mmHg), or equivalent
Setup Connect the test cables to IBP patient simulator.
Do This... Observe This... Pass/Fail
1. Connect the IBP patient simulator to the
Propaq P1 channel.
2. Set the IBP patient simulator to 0 mmHg.
3. On the Propaq MD, press > IBP.
4. On the Propaq MD, Press > P1->0<-. Display of 0/0(+/-2) in the P1 box
at the bottom of the display (P1 is
zeroed).
The waveform for channel 1 is
present.
oo
oo
5. Set the IBP patient simulator to 200
mmHg (or as close to 200 mmHg as your
simulator allows).
Display of 200/200(+/-5) in the P1
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P1
diastolic pressure is 200 (+/-5)).
oo
6. Repeat steps 1 through 5 of this procedure
for P2 and P3 channels.