Service manual

CHAPTER 1MAINTENANCE TESTS
32 9650-0801-01 Rev. A
EtCO
2
Test
Tools Needed CO
2
sampling line airway adapter
Test Setup None
Do this... Observe this... Pass/Fail
1. On Propaq MD, install the airway adapter
set and sampling line in the connector at
the side of the unit.
2.
Press the
CO
2
quick access key () to
activate the CO
2
trace.
The CO
2
trace is on the display.
oo
3. On Propaq MD, gently blow into the
airway adapter set while observing the
display.
The CO
2
waveform changes on the
display.
oo
CO
2
CO
2