Service manual

CHAPTER 1MAINTENANCE TESTS
28 9650-0801-01 Rev. A
Synchronized Cardioversion Test
Tools Needed Fluke Impulse 7000DP or equivalent defibrillator analyzer.
Test Setup Connect the cable to the defibrillator analyzer.
Select cardioversion on analyzer. Input 1mV ECG signal at 60 -120 BPM.
Do this... Observe this... Pass/Fail
1. Press the power switch to turn on the
Propaq MD, and press the Leads quick
access key
() to select PADS.
2. Press the SYNC softkey on the
defibrillator. Enter synchronized
cardioversion timing test mode on the
defibrillator analyzer.
Sync appears on display.
Sync markers display on the monitor.
The sync marker appears as an S over
the ECG R-wave peaks on strip chart
and display.
oo
3. Select 200J.
4. Press the CHARGE button. When the
SHOCK button lights, press and hold
the SHOCK button.
Observe that the R-wave to shock delay
(sync delay) is less than 60 milliseconds
on the analyzer display.
Defibrillator discharges.
oo
I, II,
III...