CHAPTER 1MAINTENANCE TESTS
22 9650-0801-01 Rev. A
Leads Test
Tools Needed 3-lead, 5-lead, and 12-lead cables
Test each cable separately.
Test Setup Connect the lead wires appropriate to each test to the Fluke Impulse 7000DP Analyzer or
equivalent.
Do this... Observe this... Pass/Fail
1. Press the power switch to turn on the
Propaq MD.
No ECG LEAD FAULT message
displayed.
oo
2. Disconnect one lead from the
simulator.
The ECG LEAD FAULT message displays
within 3 seconds (if configured).
oo
3. Reconnect the lead. Repeat step 2
with the remaining leads.
Wait for ECG LEAD FAULT message to
clear from the display (if configured).
oo
4. Repeat steps 2 and 3 for 5 lead and 12
lead cables.
Note: If heart rate alarm sounds, on the
Propaq MD:
• Press the Next quick access key
( ) key to access the second
set of quick access keys.
• Press the Alarm quick access key
().
• Press the Alarm Suspend key
().
oo