Service manual
Functional Verification Tests
9650-0801-01 Rev. A ZOLL Propaq MD Service Manual 21
Functional Verification Tests
Before You Begin the Functional Verification Tests
• Assemble the tools listed in following section, “Equipment You Need to Perform the
Functional Verification Tests.”
• Keep an extra fully charged Propaq MD battery available.
• Schedule an hour to conduct the entire maintenance test.
• Perform the tests in the order presented.
• Perform all the steps of each test procedure.
• Complete all the steps of the procedure before evaluating the test results.
Equipment You Need to Perform the Functional Verification Tests
For testing purposes, you can substitute an equivalent device.
• Fluke Impulse 7000DP Analyzer, or equivalent, with appropriate interface adapter
(Dynatech Nevada Inc. part number 3010-0378)
• IBP/Temp Simulator, or equivalent
• NIBP Simulator
• Single or dual lumen NIBP hose
• Adult NIBP cuff and a rigid cylinder to wrap it around
• 5% CO
2
calibration gas cylinder with CO
2
sample line (kit available from Scott Medical
Products, PN 4653ORF-BD)
• ECG Simulator; 12 Lead Simulator for 12 Lead test (e.g., Symbio CS1201)
Accessories You Need for the Propaq MD Options Functional Verification
Tests
• 3-lead, 5-lead and 12-lead ECG cables. (12-lead cable needed if 12-lead option is
installed.)
• NIBP cuff and hose
• IBP cable
• SpO
2
cable
and sensor
• EtCO
2
cable
and sensor
• Paddles
• Printer paper
• Battery
• Auxiliary power source