Service manual
Service Manual Repair Procedures 57
3. Spread the casing open to an angle of about 80 degrees.
4. Detach the following cables from the front chassis (indicated in figure below):
• Cable to P5 on the Analog board, from the left side panel ECG connector on the
rear chassis.
• Cable to P6 on the Analog board, from the left side panel IPB connector on the
rear chassis.
• Cable to P7 on the Analog board, from the left side panel temperature jack on the
rear chassis.
• Cable to P8 on the lower (Digital) board, from the right side panel on the rear
chassis.
• Cable to P9 on the Digital board, from the right side panel on the rear chassis.
DISCONNECT 2
UNDERLYING
DIGITAL BOARD
DISCONNECT
3 CABLES
CABLES FROM