Service manual
54 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor
2. Use the flat blade screwdriver to gently loosen the expansion module from the
monitor (see inset above).
3. Slide the monitor forward until you can disconnect the battery pack cable from the
monitor's power cable. The battery pack should remain in the expansion module
chassis.
4. Continue to slide the monitor forward until it can be completely separated from the
expansion module.
5. Slide the dual battery pack forward out of the expansion module until the battery
connector can be reached.
6. Disconnect the battery connector from the expansion module (see figure below).