38 Calibration Welch Allyn Propaq Encore Vital Signs Monitor
5. Reposition the tube to view the brightest area (now dim) and simultaneously a lighted
area (pixels on). Adjust RP7 clockwise until the background area just begins to
illuminate.
Main Power Supply Adjustments
The following procedures assume you are continuing from the recharger board calibration.
If not, refer to the setup procedures provided at the beginning of this section.
1. Turn off the monitor.
2. Remove the metal shield from the analog board. (Note the two different screw sizes;
refer to “Repair Procedures” on page 47 for instructions, if necessary.)
3. Connect the DMM reference lead to TP39.
4. Connect the DMM positive lead to TP29.
5. Turn on the monitor.
Press any of the lower five buttons to reset the Encore following equipment alerts.
6. Adjust RP7 to 5.000 V ± 0.010 V dc.
7. Refer to the table below and check the dc supply voltages using the DMM.
Table 11. Main Power Supply Voltage Checks
Supply Voltages Ref. Lead Pos. Lead Voltage Limits
Non-isolated
+15 TP39 TP3 +14.7 to +15.3
–15 TP39 TP11 –14.8 to -15.5
Isolated
+VR TP43 or TP1 TP9 +4.91 to +5.09
+VS TP43 or TP1 TP5 +6.58 to +7.15
–VS TP43 or TP1 TP21 -6.65 to -7.15