Service manual

Service Manual Functional Verification 17
Sequence for All Other Software Versions
1. Press ECG/RESP > MORE. Then press CHANGE to select the four HR/PR TONE
levels. Verify that the tone volume changes and goes off when turned OFF. Set the
volume to LOW.
2. Press MAIN MENU.
3. Software versions 2.5X: Press SETUP > MORE > MORE > SERVICE > YES > MORE
> MORE > SETTINGS > ALL ALRM. Verify alarm status displays "TURNING OFF ALL
AUDIBLE VITAL SIGNS ALARMS", "ARE YOU SURE?", YES or NO. Then press MAIN
MENU.
4. All software versions: Press SETUP > MORE > MORE (> TIME/DAY for monitor
software less than version 2.0) and check that the displayed time and date are correct.
If incorrect, press NEXT and UP or DOWN to select and change the time and date,
and then set by pressing the ENTER button.
5. Turn off the monitor.
ECG/RESP Channel and Alarm Indicators
The following procedures verify the ECG channel and the alarm indicator drivers. The
setup procedure below is used for both the visual and the electrical checks that follow.
Setup
1. Turn on the monitor and verify that it powers up correctly.
2. Put the monitor into FACTORY ADULT MODE.
Software version 2.XX: SETUP > MORE > CHANGE > SETUP > USE NOW > YES
Software version 1.XX: SETUP > MORE > CHANGE > ADULT > YES > MORE >
PROGRAM > DEFAULT > YES
3. Press MAIN MENU.
Press Result
FREEZE/UNFREEZE Freezes the waveforms
FREEZE/UNFREEZE Unfreezes the waveforms
SETUP Changes the menu
ALARMS Changes the display to ALARM STATUS
SUSPEND (on menu) SUSPEND changes to RESUME
ALARM SUSPEND/RESUME (button) Changes RESUME to SUSPEND
ALL ALRM Changes display: Turn All Off, Are You Sure, Y or N
LIMITS Changes display to HR Limits
NEXT Moves cursor block: Next block within selected parameter limits, NEXT
PAGE to next parameter
MAIN MENU (button) Returns to the main menu
NIBP START/STOP (button) Starts the NIBP pump
NIBP START/STOP (button) Stops the NIBP pump