Service manual

Service Manual Technical Overview 115
applications, a sample of exhaled gases is obtained through a small-diameter sampling
tube and routed to the bench located on the Stick board.
Unless otherwise noted, the following details are common for bench operation in both
sampling methods.
The bench consists of infrared (IR) light source and filter, a revolving chopper wheel, and a
photodetector.
The chopper wheel is an aluminum disk containing a sealed CO
2
reference cell, an open
slot, and a light-blocking area. The chopper wheel spins in the direct path of the IR light
source.
The IR source is located on one side of the bench and is projected through the gas sample
(in the airway adapter in MCO
2
; routed directly to the bench in SCO
2
) to a photodetector
at the other side of the bench. As the IR beam passes through the gas sample to the
photodetector, the components of the spinning chopper wheel are passed through the
beam with each revolution.
The amplified photodetector output signal is modulated by the revolving chopper wheel
and is referred to as VSIG. Each rotation of the chopper wheel generates two VSIG peaks
of differing size, and a low flat spot. The voltages representing these points are called
Vpeakhi, Vpeaklo, and Vdark. These correspond to the open slot, the CO
2
reference cell
and the light blocking areas on the chopper wheel, respectively. The VSIG output from
each complete rotation of the bench chopper wheel, occurring every thirty milliseconds,
together with a known bench temperature and calibration constants stored in an
EEPROM (located in the mainstream sensor connector or the sidestream Stick PCB),
provides all required information to determine the CO
2
value.