Propaq Encore Vital Signs Monitor ® Service Manual
ii Welch Allyn Propaq Encore Vital Signs Monitor © 2011 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn.
iii Contents 1 - Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 2 - Overview . . . . . . .
iv Contents Welch Allyn Propaq CS Vital Signs Monitor 6 - Repair Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Propaq Encore Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 1 Safety Summary Warnings This Safety Summary should be read by all Propaq CS monitor users. Specific warnings and cautions are placed throughout the documentation where they apply. Caution: Federal (U.S.A.) law restricts this device to sale, distribution, or use by or on the order of a licensed medical practitioner. WARNING Place the Propaq monitor and accessories in locations where they cannot harm the patient if they fall from their shelf or mount.
2 Safety Summary Welch Allyn Propaq Encore Vital Signs Monitor WARNING To ensure patient safety, the conductive parts of the ECG electrodes (including associated connectors) and other patient-applied parts should not contact other conductive parts, including earth ground, at any time. WARNING Do not operate this product in the presence of flammable anesthetics or other flammable substance in combination with air, oxygenenriched environments, or nitrous oxide; explosion can result.
Service Manual Safety Summary 3 Definitions Note Identifies information that may be important to the reader. Caution Identifies conditions or practices that could result in damage to equipment or other property. WARNING Identifies conditions or practices that could result in personal injury. Symbols The following symbols appear in the monitor documentation and on monitor labels.
4 Safety Summary Welch Allyn Propaq Encore Vital Signs Monitor The CE mark and notified body registration number signify Propaq CS Series monitors have met all essential requirements of European medical device directives 93/42/ EEC. NRTL/C Evaluated to CSA 601-1 and UL2601-1 The Canadian Standards Association has evaluated this device according to CSA 6011 and Underwriters Laboratory Standard UL 2601-1.1 Single-use only (not reusable).
5 2 Overview Purpose and Scope The Propaq Encore Service Manual is intended as a reference for monitor maintenance and repair to the field replaceable unit (FRU) level (“Field Replaceable Units (FRUs)” on page 117). This manual provides the technically qualified service person with troubleshooting information, repair procedures, and calibration and performance verification instructions. A technical overview of the monitor subsystems is provided as an introduction to the device’s circuitry and pneumatics.
6 Overview Welch Allyn Propaq Encore Vital Signs Monitor Note Save the shipping carton and packing material for repacking the Encore monitor in case it needs to be sent to a repair center or back to Welch Allyn for service. Warranty Service DO NOT ATTEMPT TO REPAIR the Encore monitor yourself during the warranty period. For service and repair, contact Welch Allyn as described below.
Service Manual Overview 7 WARNING If the monitor is opened for repair or calibration, a dielectric strength test must be completed to ensure the integrity of the patient isolation barrier. The dielectric strength test procedure is provided in “Functional Verification” on page 11. Identifying Propaq Encore Configurations The following table identifies Propaq monitor configurations and how they are indicated. The model-option number and serial number are located on the back of the housing.
8 Overview Welch Allyn Propaq Encore Vital Signs Monitor Table 2.
Service Manual Overview 9 Self Test and Service Menus The Encore monitor includes built-in self-tests that verify the integrity of its hardware and software each time you turn it on. If the monitor turns on and displays the Main Menu, all self-tests were passed. If a self-test results in a failure, a message and error code are displayed indicating the failure mode. Refer to “Troubleshooting” on page 43 and contact Welch Allyn Technical Support for help.
10 Overview Welch Allyn Propaq Encore Vital Signs Monitor
11 3 Functional Verification Introduction The functional verification procedures ensure proper operation of the Encore monitor and its options. These procedures should be performed following module-level repairs, calibration, or whenever there is a question about the accuracy or safety of the patient functions. Use the calibration procedures (“Calibration” on page 33) to return the monitor to operation within its performance specifications.
12 Functional Verification Welch Allyn Propaq Encore Vital Signs Monitor Required Equipment You will need the following equipment to verify that all functions of the Encore operate correctly. All test equipment must be calibrated to function within parameters specified by the manufacturer, and must be traceable to a national calibration standard. Some equipment can be manufactured. See “Manufacturable Test Equipment” on page 123 for information on manufacturable test equipment. Table 3.
Service Manual Functional Verification Note 13 A hi-pot test is only required if the monitor has been opened. Risk (Leakage) Current Test A risk (leakage) current test must be performed to verify that the patient remains electrically isolated from the power circuits of the Encore. Check leakage currents using a Dynatech/Nevada 431F-1D safety analyzer or its equivalent. The source current should not exceed 10A rms.
14 Functional Verification Welch Allyn Propaq Encore Vital Signs Monitor Table 5. Summary of Connection Points and High-Voltage Test Potentials Tester's high-voltage terminal connection Tester's grounded return connection High-voltage test parameters All ECG leads on patient cable connected together DC input connector; all pins connected Output: 4000 V ac rms together Current Limit: 2.5mA Ramp-up: 8 seconds Dwell: 1 second 2.
Service Manual Functional Verification 15 4. Turn on the power adapter’s power switch. 5. Check that the green LED on the power adapter turns on and that the green LED charging indicator on the monitor’s right side panel turns on. 6. Disconnect the power adapter from the monitor. Check that the monitor’s green LED charging indicator on the right side panel turns off. Caution In the next steps, carefully check for the proper polarity of the connection between the power supply and the monitor.
16 Functional Verification Welch Allyn Propaq Encore Vital Signs Monitor 13. Vary the power supply from 12 V to 28 V and verify that the charging LED stays on. 14. Turn off the power supply. 15. Disconnect the supply from the monitor. System Tests The following procedures check that the buttons operate properly, that the display works correctly, and that the date is correctly displayed. 1. Turn on the monitor. 2. Verify that no error messages appear and the monitor correctly powers up. 3.
Service Manual Functional Verification 17 Sequence for All Other Software Versions Press Result FREEZE/UNFREEZE Freezes the waveforms FREEZE/UNFREEZE Unfreezes the waveforms SETUP Changes the menu ALARMS Changes the display to ALARM STATUS SUSPEND (on menu) SUSPEND changes to RESUME ALARM SUSPEND/RESUME (button) Changes RESUME to SUSPEND ALL ALRM Changes display: Turn All Off, Are You Sure, Y or N LIMITS Changes display to HR Limits NEXT Moves cursor block: Next block within selected p
18 Functional Verification Welch Allyn Propaq Encore Vital Signs Monitor 4. Set the patient simulator as follows: • ECG Waveform: normal sinus rhythm • ECG Rate: 80 beats per minute (bpm) • ECG Size: 1 mV amplitude • Resp Lead: Lead I • Resp Baseline Impedance: 1 K • Resp Rate: 40 breaths/minute (br/m) • Resp Breath Amplitude: 1.0 5.
Service Manual Functional Verification 19 8. Check that an equipment alarm occurs. The EQUIPMENT ALERT / MULTIPLE message should accurately indicate the removed lead. The ALARM(S) OFF light will flash. • Software version 2.5X: The ECG FAULT will flash in the upper left-hand corner of the screen. • All other software versions: The ECG FAULT and RESP FAULT will alternate in the upper left-hand corner. 9. Press any button to acknowledge. Check that the ALARM(S) OFF light is on and no longer flashing.
20 Functional Verification Welch Allyn Propaq Encore Vital Signs Monitor 27. Check that the ALARM light is OFF. 28. Press MAIN MENU. Part 2— Electrical Check Use the same setup for this check described in “Setup” on page 17. 1. Set the patient simulator to provide pacer signals. 2. If the pacer indicator is not on, press ECG/RESP > MORE and select and change the pacer setting. 3. Check that a dashed vertical line is displayed on the ECG waveform each time a pacer pulse occurs. 4.
Service Manual Functional Verification 21 11. Disconnect the scope probe and ground clip and remove the wires from the connectors. 12. Using either a LifePak 5 or LifePak 6s and appropriate Welch Allyn adapter, set up the monitor and defibrillator according to instructions provided in the Defib Sync User’s Guide. 13. Check the monitor display for sync markers on the ECG waveform. 14. Disconnect the defibrillator from the monitor.
22 Functional Verification Welch Allyn Propaq Encore Vital Signs Monitor 10. After allowing a few seconds for settling, check that the mean pressure reading is 200 mmHg ±2 mmHg. 11. Remove the cable from the P1 input jack and verify that the message: P1 FAULT TRANSDUCER NOT DETECTED appears. 12. Press any button to acknowledge the discontinuance of IBP monitoring. 13. Press MAIN MENU. NIBP 1. Connect the cuff, digital pressure meter, bulb, and the Encore together with the cuff calibration kit.
Service Manual Functional Verification Pressure (mmHg) Tolerance (mmHg) 200 ±2 100 ±2 50 ±2 20 ±2 23 8. Reduce the pressure to 0 mmHg and check that the displayed readings (PR1 and PR2) are 0 ± 2 mmHg. 9. Close the bulb valve. 10. Press CANCEL. (If there is no CANCEL button at this time, go to the next step.) 11. Disconnect the adult cuff from the test setup and connect the neonatal cuff #3 that is wrapped around a solid 3 cm diameter cylinder, the black PVC tube volume. 12. Press PUMP TEST.
24 Functional Verification 1. Welch Allyn Propaq Encore Vital Signs Monitor • Compatible adult SpO2 sensor for finger. • BIO-TEK Index 2 Series SpO2 Simulator with optical finger for sensor connection (or equivalent). Turn on power to the SpO2 simulator. 2. Connect the SpO2 cable to the monitor and the SpO2 sensor. 3. Apply the SpO2 sensor to the SpO2 simulator optical finger. 4. Set the simulator manufacturer type (“MAKE”) to Masimo. 5.
Service Manual Functional Verification 25 Sensor Compatibility Some sensors are not compatible with the particular Nellcor processing board installed in the Propaq. To determine which pulse oximetry optional processing board is installed in the Propaq, find the option number on the back label of the Propaq, then refer to the table “Expansion Module Option Configurations” on page 8. The following table shows how to identify compatible sensors for each Nellcor option: Table 6.
26 Functional Verification Welch Allyn Propaq Encore Vital Signs Monitor 15. Press the START/STOP button. 16. After the printer runs out of paper, check that an equipment alarm and PAPER OUT message appear on the monitor screen. 17. Open the printer door and reload the paper into the printer. 18. Close the paper door and set the monitor upright. 19. Simultaneously press the PRINT TRENDS and START/STOP buttons. 20. Check that a test printout similar to one shown in the figure below is printed.
Service Manual Functional Verification 27 Local Barometric Pressure = 0.033421 x LUP x ATIS (ATIS in inches of Hg) 0.00131579 x LUP x ATIS (ATIS in mm of Hg) Example 1 a. The altitude is 5000 feet. The local uncorrected pressure (LUP) is determined from the following table to be 632.35. b. The reported ATIS is 30.09 inches Hg. (Altimeter: three zero zero niner.) c. Multiply the LUP (632.35) times the ATIS (30.09) times 0.033421. This gives the local barometric pressure of 635.92 mmHg. Example 2 a.
28 Functional Verification Welch Allyn Propaq Encore Vital Signs Monitor Table 7. Local Uncorrected Barometric Pressure (LUP) as a Function of Altitude (Feet) Altitude (Feet) -600 -500 -400 -300 -200 -100 0 100 200 300 400 500 600 700 800 900 1000 Pressure (mmHg) 776.62 773.83 771.05 768.28 765.51 762.75 760.00 757.26 754.52 751.80 749.08 746.37 743.67 740.97 738.28 735.61 732.93 Altitude (Feet) 2400 2500 2600 2700 2800 2900 3000 3100 3200 3300 3400 3500 3600 3700 3800 3900 4000 Pressure (mmHg) 696.
Service Manual Functional Verification 29 Table 8. Local Uncorrected Barometric Pressure (LUP) as a Function of Altitude (Meters) Altitude (Meters) 1000 Pressure (mmHg) 674.11 Altitude (Meters) 2700 Pressure (mmHg) 546.23 Altitude (Meters) 4400 Pressure (mmHg) 438.74 Altitude (Meters) 6100 Pressure (mmHg) 349.05 Mainstream CO2 Verification Procedure This procedure verifies the accuracy of the Encore’s mainstream CO2 (MCO2) measuring system.
30 Functional Verification Welch Allyn Propaq Encore Vital Signs Monitor 12. Disconnect the CO2 sensor from the monitor. 13. Turn off the monitor. 14. Log the test date and results in your department log. WARNING Eliminate all the possibilities causing errors shown in the following table before performing the calibration procedure (“MSP Board (MCO2) Calibration” on page 42).
Service Manual Functional Verification 31 6. Verify that BARO TRUE and INT are within 15mmHg (5mmHg, typical) of the barometric pressure determined in step 4. 7. Connect the test tubing setup shown in the figure below to both ends of a test water trap. 300 mmHg Manometer Test Watertrap mmHg Luer Connectors Pressure Meter Calibration Adapter Kit Standard adult cuff wrapped around arm-sized cylinder. 1097-17 8.
32 Functional Verification Welch Allyn Propaq Encore Vital Signs Monitor Caution Do not exceed 100mmHg during the following pressurization. Excessive pressure can open internal tubing connections and cause leakage. 15. Use the pump bulb to manually inflate the pressure to 100mmHg. 16. Wait 15 seconds for the pressure to stabilize. Note the pressure reading, then wait an additional 60 seconds and record the leak rate. This is the total leakage (including water trap leakage from step 9). 17.
33 4 Calibration Introduction This section provides calibration procedures for the Propaq Encore. These procedures allow you to calibrate the Encore to Welch Allyn factory specifications. Calibrations procedures should be performed only by qualified service personnel. Calibration procedures must be performed whenever a question arises as to the performance of the Encore, and as indicated in “Overview” on page 5.
34 Calibration Welch Allyn Propaq Encore Vital Signs Monitor Table 10. Required Equipment Description Adult cuff (provided with monitor); cylindrical object about the size of an adult’s arm Welch Allyn cuff calibration kit, Welch Allyn part number 008-0012-XX Propaq Encore ac power adapter (provided with monitor) Resistor test loads (4.9¾ and 4.65¾). Refer to “Manufacturable Test Equipment” on page 123.
Service Manual Calibration Note 35 To disable the alarms while performing the calibration procedures, press SETUP > ALL ALRM ( >YES - version 2.0 software)to remove the bell symbol from the display. In order to inhibit the Apnea alarm, turn off Respiration monitoring: press ECG/RESP > MORE and set RESP to OFF. Recharger Supply Adjustments WARNING High voltages (approximately ± 200V dc or 360V ac) are present on the recharger board when the Encore is turned on.
36 Calibration Welch Allyn Propaq Encore Vital Signs Monitor 0696-02 RIBBED (+) 7. SMOOTH (-) Connect the adapter cable to the monitor’s dc input jack. 8. Verify that the input current from the variable dc power supply is less than 1.8A and that the green LED on the Encore’s right side panel is on. Note Refer to the figure below for adjustment and test point locations. 9. Connect the DMM positive lead to P12 pin 1 or 5. 10. Connect the DMM reference (negative) lead to non-isolated ground at TP2.
Service Manual Calibration 37 Display Voltage Adjustments Note Do not adjust the display voltage unless the display has been changed and the lack of contrast between illuminated and non-illuminated pixels affects the display readability. The following calibration adjusts the monitor display’s high voltage supply which is located on the recharger board. The factory-set voltage is specific to the display and is indicated on the display module label.
38 Calibration Welch Allyn Propaq Encore Vital Signs Monitor 5. Reposition the tube to view the brightest area (now dim) and simultaneously a lighted area (pixels on). Adjust RP7 clockwise until the background area just begins to illuminate. Main Power Supply Adjustments The following procedures assume you are continuing from the recharger board calibration. If not, refer to the setup procedures provided at the beginning of this section. 1. Turn off the monitor. 2.
Service Manual Calibration 39 Table 11. Main Power Supply Voltage Checks Supply Voltages Ref. Lead Pos. Lead Voltage Limits +VB TP43 or TP1 TP13 +5.40 to +6.10 +5.6 TP43 or TP1 TP33 +5.55 to +5.70 TP43 or TP1 P6, Pin 10 +2.47 to +2.53 TP43 or TP1 U5, Pin 1 0.00 to +0.25 NIBP Pressure Offset TP43 or TP1 TP31 +2.40 to +2.60 OSC baseline TP43 or TP1 TP47 -0.95 to -1.25 +VA TP43 or TP1 TP7 (+VR + 0.1V) to (+VR + 0.2V) -VA TP43 or TP1 TP15 (+VR + 0.2V) to -(VR + 0.35V) +2.
40 Calibration Welch Allyn Propaq Encore Vital Signs Monitor Non-Invasive Blood Pressure Calibration 1. Connect the cuff, digital pressure meter, bulb, and the Encore together with the cuff calibration kit. Using T-connectors from the kit, connect the cuff, pressure meter, and bulb. 2. Connect this NIBP test setup to the pressure transducer at PT1 on the Analog board. 3. Wrap the adult cuff around the black tube volume or a cylindrical object about the size of an adult’s arm. 4.
Service Manual Calibration 41 11. Connect the oscilloscope reference lead to TP43 or TP1 on the Analog board. 12. Check for 15mV peak-to-peak noise. Note Measure the low frequency noise (bounce) only. Make the peak-to-peak measurement from the center-to-center of the trace to disregard the “fuzzy” high frequency noise. 13. Disconnect all patient inputs to the Encore’s left side panel. EMI Null Adjustment 1. Touch the oscilloscope probe to TP45 on the Analog board (square metal pad). 2.
42 Calibration Welch Allyn Propaq Encore Vital Signs Monitor 9. Set the simulator to 200mmHg. 10. Check that P2 displays 200mmHg ±2 mmHg. 11. Disconnect all patient cables from the left side panel. 12. Reinstall the Analog board shield and torque it according to the procedure in “Repair Procedures” on page 47. MSP Board (MCO2) Calibration The Propaq MSP board has two calibration adjustments.
43 5 Troubleshooting Introduction This section provides information that can be helpful in troubleshooting the Propaq Encore monitor. Screen Messages Messages may appear on the display to inform the operator of some condition requiring operator or service attention. Messages that indicate that the Propaq Encore may need servicing are listed below. For information about other messages, refer to the Propaq Encore Reference Guide. PROGRAM FAULT: SETTINGS LOST, TIME/DAY RESET.
44 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor LACK OF WAVEFORM, SENSOR DISABLED - (MCO2). The airway adapter (cuvette) may be contaminated or the light source in the sensor may be faulty. Clean or replace the cuvette as required or replace the CO2 sensor. SENSOR FAILURE, CALIBRATION ERROR - (MCO2). A Sensor failure has occurred. Replace the sensor. PUMP FAILURE, SERVICE REQUIRED - (SCO2). The sidestream pneumatic tubing is kinked or a pump failure on the SSP board has occurred.
Service Manual Troubleshooting x=7 indicates a problem found by the Side Stream Processor (SSP) on the sidestream main board. yyy-zzzz indicate specific types of problems. Contact Welch Allyn Technical Support for more details. 45 Battery Capacity Check Several variables affect monitor operating time on battery: • active options • frequency of NIBP measurements • frequency and length of printouts • ambient temperature, • battery age and condition • displayed information.
46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor
47 6 Repair Procedures Introduction This section provides instructions for removing and replacing serviceable modules in the: • Propaq Encore monitor • expansion module (which may include any or all of the following: printer, SpO2, mainstream CO2 ,sidestream CO2) • SpO2-only module Instructions on how to remove the battery pack are followed by instructions on how to disassemble the monitor, the expansion module, the printer, the SpO2 module, and the CO2 module.
48 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor WARNING STATIC DISCHARGE All calibration, service, and repair must be done at a static-protected work station. Failure to protect against damage due to static discharge may reduce monitor operation life and increase repairs. Place the monitor in an anti-static area before servicing the monitor. Required Tools The following tools are required to disassemble and reassemble the Propaq Encore components.
Service Manual Repair Procedures Monitor Only Monitor with SpO2 Monitor with Expansion Module 0696-10 Field Replaceable Units in the Monitor The front chassis of the monitor contains the following components: • insulator • Analog board with metal shield and insulator • Digital board with shields • EL display unit • front chassis with display filter and gasket 49
50 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor Plastic Insulator Analog Board with Shield Digital Board Top Shield Digital Board Brackets EL Display Panel Front Chassis 0696-11 The rear chassis contains the following components: • Recharger board • Side panels (right and left) • NIBP pump • NIBP tubing • NIBP valves
Service Manual Repair Procedures 51 Recharger Board Right Side Panel Left Side Panel Rear Chassis 0696-12 Replacing the Power Input Fuse The F1 fuse protects the Recharger board circuitry against excessive current at the dc input connector. You do not need to disassemble the Propaq Encore to replace this fuse.
52 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor To remove the fuse, use a flat blade screwdriver to turn the fuse cover and release the fuse. Caution Replace the fuse only with one of the same rating and size. See “Field Replaceable Units (FRUs)” on page 117 for Welch Allyn part numbers.
Service Manual Repair Procedures 53 single battery cover removal Battery Pack Battery Pack Cable Battery Cover 0696-13 2. Disconnect the battery pack cable from the battery pack. 3. Remove the battery pack from the monitor. 4. Store the battery in a safe place while disassembling the monitor. Removing the Dual Battery in Expansion Module The expansion module uses a dual battery pack. The following instructions describe how to remove the dual battery. Note 1.
54 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor 2. Use the flat blade screwdriver to gently loosen the expansion module from the monitor (see inset above). 3. Slide the monitor forward until you can disconnect the battery pack cable from the monitor's power cable. The battery pack should remain in the expansion module chassis. 4. Continue to slide the monitor forward until it can be completely separated from the expansion module. 5.
Service Manual Repair Procedures 7. 55 Remove the dual battery pack from the expansion module. 8. Store the battery in a safe place while disassembling the monitor and expansion module. Removing the Dual Battery in SpO2 Module The steps for removing the battery pack from the SpO2 module are identical to those in removing the battery from the expansion module, except you remove all screws on the rear panel. The monitor is secured to the SpO2 module with four screws; each screw has two washers. 1.
56 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor 0696-18 Caution Before opening the casing more than one inch in the next step, disconnect the tube from the pressure transducer at the joint. Failure to disconnect the tubing may cause damage to the transducer or tubing. 3. Slightly spread the casing open at the top until you can disconnect the pressure transducer tube indicated in the figure below. 1. Spread the casing open to an angle of 45 degrees. 2.
Service Manual Repair Procedures 57 3. Spread the casing open to an angle of about 80 degrees. 4. Detach the following cables from the front chassis (indicated in figure below): • Cable to P5 on the Analog board, from the left side panel ECG connector on the rear chassis. • Cable to P6 on the Analog board, from the left side panel IPB connector on the rear chassis. • Cable to P7 on the Analog board, from the left side panel temperature jack on the rear chassis.
58 Repair Procedures 1. Welch Allyn Propaq Encore Vital Signs Monitor Detach the two ribbon cables from the rear chassis: • Ribbon cable to P10 on the Recharger board, from the Digital board. • Ribbon cable to P11 on the Recharger board, from the Digital board. Note The two halves are sealed by rubber O-ring gaskets that are recessed around the perimeter of the rear chassis. If the gaskets become loose, replace them in the perimeter before re-assembling the monitor. 2.
Service Manual Repair Procedures 59 Closing the Monitor 1. Before closing the monitor casing, verify the following: • Cable to P6 on the Analog board goes behind and below the cable to P5, and over the cable to P7. • Cable to P5 on the Analog board goes over and in front of the cable to P6. • Cable to P7 on the Analog board goes under the cable to P6. • Cable to P1 (the expansion cable) must not lay over the transformer.
60 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor Note 1. You do not need to lift the plastic insulator nor remove the metal shield to remove the Analog board. However, you must lift the insulator to calibrate the board. Remove the battery pack (“Replacing the Battery Pack” on page 52). 2. Open the monitor (“Opening the Monitor” on page 55). 3. Remove the five screws fastening the Analog board to the front chassis.
Service Manual Repair Procedures 61 When installing the Analog board: • Blow out any dust and particulates from the board using pressurized de-ionized gas before installing it. • Align the board so that P4 correctly plugs into J4. • Torque the screws holding the Analog board to 3.8in-lbs to 4.2in-lbs (.43N-m to .47N-m). Replacing the Digital Board If a problem occurs with the Digital board, return the entire monitor to Welch Allyn for service. See “Technical Support Services” on page 6.
62 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor 5. Use the nut driver to loosen and remove the hex post. 6. Lift the Digital board up and out of the front chassis. Note 7. The Digital board is connected to the EL display module by connector J1. This connector should detach easily. Remove the 10-pin ribbon cable (cable from the Digital board to the Recharger board). Caution Do not touch the button contact surfaces on the underside of the Digital board.
Service Manual Repair Procedures When installing the Digital board: • Make sure the cable between the Digital board and the Recharger board is installed on the Digital board before placing the board in the front chassis. • Remove the shield on the top of the Digital board in order to reconnect the display cable.
64 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor • Replace the shield, noting the correct screw lengths indicated on the shield. When replacing the insulating shield, make sure the left side of the shield is placed underneath the top surface of the bracket. • Torque all screws and the hex post to 3.8in-lb to 4.0in-lb (0.43 N-m to 0.47 N-m).
Service Manual Repair Procedures REMOVE 4 SCREWS AT CORNERS 65 HOLD THE GASKET DOWN AT THESE POINTS ON BOTH SIDES OF THE BOARD. Caution In the next step, if the gasket lifts off the display filter at any point, you must inspect the display filter to verify that the conductive coating on the filter is still intact. See the steps below. 6. To carefully remove the display: a.
66 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor OTI Coating Approx. 0.125" 0696-33 When installing the display module: • Remove any dust from the display filter using pressurized de-ionized gas. • Orient the display module so that the J1 connector is toward the buttons. • Orient the brackets so that the holes on the bracket straps align with the holes in the posts on the sides of the chassis. • Verify two washers are present for each screw. • Torque screws to 3.8in-lbs to 4.
Service Manual Repair Procedures 67 4. Remove the Digital board (“Replacing the Digital Board” on page 61). 5. Grasp the Silastic gray dome and pull the dome off the button post. 6. Squeeze the plastic button latches together and then push the button out the front of the panel. 0696-34 Note The button dome can be replaced on the post without regard to the orientation of the notch engraved in the dome.
68 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor F2 P2 T3 D33 Replacing Air Tubing Replace the NIBP tubing as a single assembly to ensure accurate tubing lengths. Refer to “Field Replaceable Units (FRUs)” on page 117 for the tubing assembly part number. Note If you don’t need to replace the length of tubing between the cuff hose connector in the left side panel and the rest of the tubing assembly, don’t remove the left side panel to disconnect this tube.
Service Manual Repair Procedures 69 Replacing Cables If you suspect a cable to be faulty, check its continuity on all pins. If you need to replace a cable, refer to “Field Replaceable Units (FRUs)” on page 117 for part numbers.. When removing a cable, be sure to note the routing. It is important that the new cable is routed in the same manner.
70 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor Replacing PROMs Read “Propaq Encore Software” on page 47. The PROMs require a special removal tool to replace them. (“Required Tools” on page 48.) To prevent damage to the PROM sockets, do not use any other tools. Caution The PROMs can be damaged by static discharge. Work only at a staticprotected work station. Do not remove the PROMs from the Analog and Digital board.
Service Manual Repair Procedures 71 Cuff Fittings Air system integrity in the NIBP system is essential for accurate measurements. Routine wear of the O-ring gasket located in the threaded fitting of the cuff hose can result in air leaks. If you encounter air leaks that cannot be found from the tests in this manual, replace the O-ring first. WARNING Use of NIBP channel with leaking cuff hose fittings may result in inaccurate noninvasive blood pressure measurements. 1.
72 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor Replacing the Recharger Board The Recharger board is secured to the rear chassis by four screws. When removing the Recharger board, note the location of cables, plumbing, and screws and screw lengths. 1. Remove the battery pack (“Replacing the Battery Pack” on page 52). 2. Open the monitor (“Opening the Monitor” on page 55). 3. Disconnect the air tubing attached to the valves and pump at the junction indicated in this figure.
Service Manual Repair Procedures Screws Screws 0696 40 6. Lift out the Recharger board. Caution As you set the board in place during installation, keep the board clear of the power switch solder terminals to avoid binding against them. The terminals may become damaged if force is applied against them. When installing the Recharger board: • blow out any dust and particulates using pressurized de-ionized gas • move all cables out of the way • torque screws to 3.8in-lbs to 4.2in-lbs (.43N-m to .
74 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor Replacing the Pump New wires and a connector are provided with the replacement pump kit. 1. Remove the battery pack (“Replacing the Battery Pack” on page 52). 2. Open the monitor (“Opening the Monitor” on page 55). 3. Remove the Recharger board (“Replacing the Recharger Board” on page 72). 4. Unplug the pump cable from the Recharger board. 5. Disconnect the tubing from the pump. 6.
Service Manual Repair Procedures 75 Replacing the Side Panels The Propaq Encore side panels are complete assemblies. The side panels are sealed to the rear chassis with an adhesive gasket. A new gasket is provided with the new side panel kit. Removing the Right Side Panel 1. Remove the battery pack (“Replacing the Battery Pack” on page 52). 2. Open the monitor (“Opening the Monitor” on page 55). 3. Remove the Recharger board (“Replacing the Recharger Board” on page 72). 4.
76 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor INSERT SMALL SCREWDRIVER BETWEEN PLASTIC AND CLAMP TO RELIEVE TENSION AGAINST PLASTIC 5. Press the side panel out of the rear chassis. There is an adhesive gasket around the side panel which requires considerable force to separate. Do not pry against components. 6. Remove the tubing attached to the cuff hose connector. You can reuse this tubing when installing the new side panel. Installing the Side Panels 1.
Service Manual Repair Procedures Note 77 Uncovering the protective paper in the next step exposes the adhesive on the gasket. Do not touch or otherwise contaminate the adhesive, or the adhesive may not properly seal the side panel. 4. On the new side panel, remove the protective paper from the gasket attached to the side panel. Note When installing the left side panel in the next step, slightly bend up wires at the P1 and P2 pressure connector pins to allow proper fit. 5. Set the side panel in place.
78 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor Expansion Module The expansion module housing contains one or all of the following: • printer mechanism with printer main board • MSP board (units with CO2 option) • CO2 connector side panel • SCP/SpO2 board (units with printer and SpO2 options) • SpO2 connector side panel • SpO2 speaker side panel • front panel assembly with pushbuttons • SSP board (units with sidestream CO2) • Stick board (units with sidestream CO2) •
Service Manual Repair Procedures 79 Opening the Expansion Module Rear Panel 1. Separate the monitor and expansion module (“Removing the Dual Battery in Expansion Module” on page 53). Note For all sidestream CO2 options (226, 227, 228, or 229) remove the front panel (“Replacing the Expansion Module Front Panel” on page 81) and the extender board “Replacing the Extender Board” on page 88 before proceeding with the rear-panel removal. 2.
80 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor Closing the Expansion Module 1. Make sure all cables are in place and the gaskets removed earlier are available. 2. Set the rubber gaskets in their proper places as shown in the figure below. 3. Connect the printer cable and expansion cable to the SCP or MSP board. Gasket Material 4. Set the chassis onto the rear panel, verifying that the gaskets don’t slip out of a channel and become pinched between the rear panel and the chassis.
Service Manual Repair Procedures 81 Replacing the Expansion Module Front Panel The front panel contains the printer buttons. Like the rear panel, the front is sealed to the chassis with a rubber O-ring gasket. This gasket can be re-used when replacing the front panel. The expansion module front panel is attached to the chassis with four screws. Use the following procedure to remove the front panel: 1. Remove the four screws (as shown below) holding the front panel. 2.
82 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor Replacing the Expansion Module Button Board and Buttons The expansion module contains a button board that has electrical contact areas to detect when an expansion module button is pressed. The front panel buttons are similar to the monitor buttons. A plastic button and flexible contact touches the contact area on the button board when a button is pressed. Use the following procedure to remove the button board and buttons: 1.
Service Manual Repair Procedures 83 Replacing the Printer This section provides instructions on removing and installing the printer mechanism and printer main board. These two components are provided as a replacement unit called the printer assembly. Caution Do not attempt to separate the printer mechanism from the printer main board. Replace these as a unit. Removing the Printer Remove the printer assembly according to the following instructions. 1. Remove any printer paper. 2.
84 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor Installing the Printer 1. Lay the chassis on its back 2. Completely open the paper door. Caution As you install the printer assembly, rotate the printer door latch mechansim out of the way so it clears the chassis. The assembly fits snugly, but should not require much force to slide it in place. 3.
Service Manual Repair Procedures 85 Opening the SpO2 Module 1. Separate the SpO2 module from the monitor (“Removing the Dual Battery in SpO2 Module” on page 55). 2. Two tabs secure the SpO2 module rear panel to the SpO2 module chassis. Using a small screwdriver or your finger, press on each of the tabs inside the SpO2 module chassis while applying pressure to separate the rear panel from the chassis, as shown here.
86 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor Replacing the SpO2 Circuit Boards This describes how to replace the two SpO2 circuit boards in the SpO2 module and in the expansion module without a CO2 option. To replace SpO2 boards in an expansion module with a CO2 option, see “Replacing the MSP/SpO2 Circuit Boards” on page 87. Caution When replacing the SpO2 circuit boards, remove both circuit boards with the patient connector side panel still connected.
Service Manual Repair Procedures 87 Replacing the MSP/SpO2 Circuit Boards This describes how to replace the MSP/SpO2 circuit boards in the expansion module with a CO2 option. To replace SpO2 boards in the SpO2 module and the expansion module without a CO2 option installed, see “Replacing the SpO2 Circuit Boards” on page 86. Caution When replacing the MSP/SpO2 circuit boards, remove both circuit boards with the patient connector side panel still connected.
88 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor • set the two circuit boards and side panels in place, making sure that the boards are seated on the support tabs • torque screws to 4.75in-lbs to 5.25in-lbs (.54N-m to .59N-m) • reconnect the mainstream CO2 patient connector cable Replacing the Sidestream CO2 Assemblies The following instructions describe how to replace the sidestream CO2 option assemblies in the expansion module.
Service Manual Repair Procedures 89 Removing the SSP/Stick board assembly: Caution When servicing the SSP and Stick board assemblies, be sure to note the tubing routing prior to disconnection. Always reassemble tubing exactly as removed. J7 Stick Board Retainer Retainer Pump Pressure Transducer No connection 1097-18 SSP Board Screws Pneumatic Reservoir J4 Water Trap Receiver Exhaust Port Hole 1097-19 1.
90 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor Removing the Stick board from the SSP board: 1. Disconnect the tubing from the pump and pressure transducer PT2 on the frontfacing side of the SSP board and the pneumatic reservoir on the rear-facing side. 2. Remove the two screws from the front-facing side of the SSP board that secure two slotted nylon Stick board retainers. 3. Pull the Stick board up to free it from connector J7 and remove it from the SSP board.
Service Manual Repair Procedures 91 Replacing Expansion Module Side Panels The side panels easily slide out after the chassis and rear panel are separated. If the patient connector side panel is being replaced, this side panel and the two circuit boards must be removed together before disconnecting the side panel from the circuit board. See “Replacing the SpO2 Circuit Boards” on page 86 or “Replacing the MSP/SpO2 Circuit Boards” on page 87.
92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor
93 7 Technical Overview This section describes the electronics of the Encore monitor and three of its options: Printer, Pulse Oximetry (SpO2), and Capnography (CO2). This section is intended as an aid to servicing at the field-replaceable unit level and contains only system-level and functional-level circuit descriptions. When replacing modules or components, refer to “Repair Procedures” on page 47. For part numbers, see “Field Replaceable Units (FRUs)” on page 117.
94 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor System Description The Encore monitor can be configured as a standalone monitor or with one of two optional housings containing different monitoring and printer options.
Service Manual Technical Overview Filler Panel Rear Panel Main Housing Filler Panel Printer Expansion Module Option 221 printer Printer Door Front Panel, no SCO2 1097-04 SpO2 Connector Panel Rear Panel SpO2 Board Speaker Panel Main Housing SCP Board Printer Printer Door Front Panel, no SCO2 Expansion Module Option 223 printer SpO2 1097-05 95
96 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor SpO2 Connector Panel Rear Panel CO2 Connector Panel SpO2 Board Speaker Panel Main Housing MSP Board Expansion Module Option 224 SpO2 MCO2 Front Panel, no SCO2 1097-06 SpO2 Connector Panel Rear Panel CO2 Connector Panel SpO2 Board Speaker Panel Main Housing 08 MSP Board Printer Printer Door Front Panel, no SCO2 Expansion Module Option 225 printer SpO2 MCO2 1097-07
Service Manual Technical Overview SpO2 Connector Panel Rear Panel SpO2 Board Speaker Panel Main Housing MSP Board Expansion Module Option 226 SpO2 SCO2 SSP Board Extender Board Front Panel, with SCO2 1097-08 Stick Board SpO2 Connector Panel Rear Panel SpO2 Board Speaker Panel Main Housing MSP Board Printer SSP Board Printer Door Extender Board Front Panel, with SCO2 Stick Board Expansion Module Option 227 printer SpO2 SCO2 1097-09 97
98 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor SpO2 Connector Panel Rear Panel CO2 Connector Panel SpO2 Board Speaker Panel Main Housing 08 MSP Board Expansion Module Option 228 SpO2 MCO2 SCO2 SSP Board Extender Board Front Panel, with SCO2 Stick Board 1097-10 SpO2 Connector Panel Rear Panel CO2 Connector Panel SpO2 Board Speaker Panel 03 Main Housing MSP Board Printer SSP Board Printer Door Extender Board Front Panel, with SCO2 Stick Board Expansion Module Option 229 pri
Service Manual Technical Overview monitor system ENCORE SYSTEM BLOCK DIAGRAM MONITOR CHASSIS ANALOG BOARD ANALOG PATIENT DATA means optional configuration (PATIENT DATA ACQUISITION, ANALOG-TO-DIGITAL CONVERSION, ISOLATED CPU, ISO/NON-ISO POWER SUPPLIES, REAL-TIME ECG OUT) REAL-TIME ECG OUT LOCATED IN FRONT CHASSIS CUFF PRESSURE DC POWER RECHARGER BOARD DISPLAY POWER/ CONTROL DATA DIGITAL BOARD (DISPLAY PS, BATT RECHARGER NIBP PUMP & VALVES) LOCATED IN REAR CHASSIS (NON-ISOLATED CPU, DISPLAY
100 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor Power Electronics The power system circuits are located on the Analog, Digital, and Recharger boards. The power system includes the ac mains power adapter.
POWER ADAPTER EL DISPLAY POWER SUPPLY +BAT POWER SWITCH +VSW VCC T1PIN9 T1PIN7 T1PIN8 +15V GND -15V PS_CONTROL PWR_SWT_COM VCC SHUTDOWN/ HVENBL/ DISPL_EN ISOLATED POWER SUPPLIES IAG ANALOG BOARD 5.9V SUPPLIES FOR ISOLATED RS423 T1 BATT_TAP LOW_VCC INFREQ_INT/ +VSW MONITOR POWER CONTROL PIN9 (ISOLATED GND) PIN7 PIN8 0696-59 (RS423 POWER) NON-ISO POWER SUPPLIES VCC SUPPLY +5.00V ISO 5.
102 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor Monitoring Electronics System The monitoring electronics reside on the Analog and Digital boards located in the front chassis assembly.
Service Manual Technical Overview 103 NIBP Pneumatics The NIBP (cuff) system: • inflates the cuff to a target cuff pressure • converts the cuff air pressure to an electrical signal using a pressure transducer • deflates the cuff in a controlled manner while the monitoring electronics determines blood pressure • protects the patient from cuff over-inflation The NIBP air system consists of the pump, air valves, transducers, and the cuff.
104 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor NIBP PNEUMATICS AT REST BETWEEN MEASUREMENTS TO DAP A-D CONVERTER PRIMARY TRANSDUCER SERVO (LINEAR) VALVE CLOSED CUFF DUMP VALVE OPEN PUMP BLEED VALVE OPEN OFF SECONDARY TRANSDUCER TO DCP A-D CONVERTER ATMOSPHERIC PRESSURE 0696-62 INCREASING PRESSURE (INFLATION) DECREASING PRESSURE (DEFLATION) PRESSURIZED (TARGET REACHED) NIBP PNEUMATICS DURING PUMP UP TO TARGET PRESSURE TO DAP A-D CONVERTER PRIMARY TRANSDUCER CUFF SERVO (LI
Service Manual Technical Overview NIBP PNEUMATICS AT MAXIMUM TARGET PRESSURE TO DAP A-D CONVERTER PRIMARY TRANSDUCER CUFF SERVO (LINEAR) VALVE CLOSED DUMP VALVE CLOSED BLEED VALVE OPEN PUMP OFF SECONDARY TRANSDUCER TO DCP A-D CONVERTER ATMOSPHERIC PRESSURE INCREASING PRESSURE (INFLATION) 0696-64 DECREASING PRESSURE (DEFLATION) PRESSURIZED (TARGET REACHED) NIBP PNEUMATICS DURING MEASUREMENT BLEED TO DAP A-D CONVERTER PRIMARY TRANSDUCER CUFF SERVO (LINEAR) VALVE PARTIALLY OPEN DUMP VALVE CLOSED B
106 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor NIBP PNEUMATICS AFTER MEASUREMENT COMPLETE TO DAP A-D CONVERTER PRIMARY TRANSDUCER CUFF SERVO (LINEAR) VALVE CLOSED* DUMP VALVE OPEN BLEED VALVE OPEN PUMP SECONDARY TRANSDUCER OFF TO DCP A-D CONVERTER ATMOSPHERIC PRESSURE INCREASING PRESSURE (INFLATION) DECREASING PRESSURE (DEFLATION) PRESSURIZED (TARGET REACHED) * VALVE IS INITIALLY CLOSED, BUT OPENS AFTER INITIAL DUMPING BEGINS BY DUMP VALVE.
Service Manual Technical Overview Cabling Diagrams cabling diag 107
108 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor Expansion Module and Printer The printer shares battery power with the monitor, but contains its own power supply. The printer processor communicates with the monitor's processor through the expansion connector and cable. The Expansion Module is designed to house the Propaq Encore Printer and other circuit boards for additional parameters. The main circuit board in the Printer is the Printer Main Board.
Service Manual Technical Overview PRINTER FUNCTIONAL BLOCK DIAGRAM P502 P150 FROM BATTERY PRINTER POWER PWR (0..9) PRINTER DRIVER TO MOTOR P501 TO PAPER OUT DET. MOT (0..5) P402 P1 EXPANSION VCC SYSTEM POWER PRINTER CPU TO HEAD UP DET. P400 TO KEY BOARD AD (0..7) P140 FROM/TO PROPAQ PRINT HEAD HEAD (0..
110 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor Pulse Oximetry (SpO2) Option The monitor pulse oximetry option (SpO2) hardware includes two circuit boards, a sidepanel connector, and related hardware. The pulse oximetry option is installed in the pulse oximetry option module attached to the rear of the monitor (also known as the Baqpaq) or the expansion module when the printer or the CO2 option are also present.
Service Manual Technical Overview 111 PULSE OXIMETRY BLOCK DIAGRAM AC COUPLED AMP INTERNAL PREAMP INVERTING AMP (VAR.
112 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor can then be digitized. The data from the ADC is processed by the CPU and then sent to the SCP board for transmission to the Propaq Encore monitor's CPU. SpO2 signal level is controlled through the variable gain stage of the inverting amplifier and the output current of the LED current driver. By controlling both signal gain and current drive, the CPU can maintain the best possible signal level for processing.
Service Manual Technical Overview 113 Capnography (CO2) Two capnography (CO2) monitoring options are available—the mainstream -sampling CO2 option (MCO2) and the sidestream-sampling CO2 option (SCO2). Both mainstream CO2 and sidestream CO2 options are available in one product. However, only one sampling method may be monitored at any time—simultaneous monitoring of MCO2 and SCO2 is not supported.
114 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor Isolated Circuitry. The Isolated Circuitry directly controls the hardware responsible for the CO2 measurement function. CPU Circuitry. The CPU Circuitry includes the Philips 80C552 micro controller, 32K of RAM, 64K of ROM, and standard RESET circuitry. A UART directly interfaces with the pulse oximetry SpO2 board providing the SpO2 measurement function.
Service Manual Technical Overview 115 applications, a sample of exhaled gases is obtained through a small-diameter sampling tube and routed to the bench located on the Stick board. Unless otherwise noted, the following details are common for bench operation in both sampling methods. The bench consists of infrared (IR) light source and filter, a revolving chopper wheel, and a photodetector.
116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor
117 8 Field Replaceable Units (FRUs) This identifies the available FRUs for the Propaq Encore. To order an FRU, contact Welch Allyn Technical Support as described in “Technical Support Services” on page 6.
118 Field Replaceable Units (FRUs) Welch Allyn Propaq Encore Vital Signs Monitor Description Part # VALVE, BLEED/DUMP (VLV1, VLV2) 680-0050-00 VALVE, LINEAR SERVO (VLV3) 680-0049-01 Miscellaneous Components for Monitor FUSE, 2AG - 3A, SLO BLO, 250 V, EXTERNAL, 1 EACH 503-0058-00 FUSES, 3A PICO FUSE, 125 V, QTY.
Service Manual Field Replaceable Units (FRUs) 119 Description Part # SPO2 CONNECTOR PANEL, NELLCOR MP-405 010-0210-02 SPO2 CONNECTOR PANEL, NELLCOR MP-507 010-0231-00 SPO2 CONNECTOR PANEL, MASIMO MS3 - MINI DIN CONNECTOR WITH SHROUD 010-0208-00 SPO2 CONNECTOR PANEL, MASIMO MS3 WITH MASIMO STYLE CONNECTOR 010-0236-00 LOCKING "D" CONNECTOR, SPO2 PANEL 630-0038-00 CO2 Option, Mainstream/Sidestream MSP BOARD, With MCO2, MP-203 ONLY, U11 64K PROM, D/E/G SERIES 020-0580-00 MSP BOARD, Without MCO2
120 Field Replaceable Units (FRUs) Welch Allyn Propaq Encore Vital Signs Monitor Description Part # LABEL, LEFT SIDE PANEL, 2X2/RESP., AAMI, UNIVERSAL 640-0329-01 LABEL, LEFT SIDE PANEL, 2X2/RESP., HP, UNIVERSAL 640-0355-01 LABEL, LEFT SIDE PANEL, 2X4, AAMI, UNIVERSAL 640-0320-01 LABEL, LEFT SIDE PANEL, 2X4, HP, UNIVERSAL 640-0354-01 LABEL, LEFT SIDE PANEL, 2X4/RESP., AAMI, UNIVERSAL 640-0328-01 LABEL, LEFT SIDE PANEL, 2X4/RESP.
Service Manual Field Replaceable Units (FRUs) 121 Description Part # LABEL, LEFT AND RIGHT KEYS WITH ALARM LEDS 640-0479-01 LABEL, LEFT AND RIGHT KEYS, NET OFF 640-0484-00 LABEL, LEFT AND RIGHT KEYS, NET OFF 640-0484-01 LABEL, PRINTER KEY PANEL 640-0372-00 Labels, German Applications LABEL, LEFT AND RIGHT KEYS WITH ALARM LEDS 640-0480-00 LABEL, LEFT AND RIGHT KEYS WITH ALARM LEDS 640-0480-01 LABEL, LEFT AND RIGHT KEYS, NET OFF 640-0483-00 LABEL, LEFT AND RIGHT KEYS, NET OFF 640-0483-01
122 Field Replaceable Units (FRUs) Welch Allyn Propaq Encore Vital Signs Monitor
123 A Manufacturable Test Equipment The following device(s) can be manufactured for use in the functional verification and calibration procedures in this manual. Parts are readily available from electronic component suppliers. (USED TO ADJUST RECHARGER POWER SUPPLIES) USE RESISTOR NETWORKS TO RIGHT AND CREATE TWO (2) SEPARATE TEST LOADS: 4.9 OHM AND 4.65 OHM LOADS 1 2 3 4 5 CONNECTOR: 5-PIN, 0.
124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor
125 B Dynatech/Nevada Patient Simulator Modification The factory configuration of the Dynatech/Nevada model 214A, 215A, and 217A Patient Simulators is incompatible with the Propaq Encore monitors with IBP channels when both the ECG and invasive pressure simulator channels are simultaneously connected to the monitor. Erroneous readings result. The patient simulator must be modified as described here before you can use both simulator channels with the Propaq Encore monitors with IBP.
126 Dynatech/Nevada Patient Simulator Modification Welch Allyn Propaq Encore Vital Signs Monitor