User Manual
Preliminary document not for redistribution. Company confidential. Product features and specifications may vary.
Copyright ©2006 Dynastream Innovations Inc. REV: 2006-08-18 Page 22
Footpod Calibration
The accuracy of the speed distance monitor (footpod) is 97% out of the box for a majority of users. Some users with
certain stride characteristics, such as pronation and supination, will not obtain 97% accuracy until a simple
calibration procedure is performed. Calibration can be done either manually or using the auto-calibration feature.
Manual Calibration
Manual CalibrationManual Calibration
Manual Calibration
• Manual calibration involves the adjustment of a multiplication factor that is then applied to all future speed
and distance measurements.
• The nominal value of the manual calibration factor is 1.000.
• The nominal value can be adjusted upwards or downwards if the athlete has a known reference distance to
calculate the calibration factor appropriately.
• The manual calibration factor will be automatically adjusted following an auto-calibration process, and can be
recorded for future reference (for example to track calibration values for multiple users).
Auto
AutoAuto
Auto-
--
-Calibration
CalibrationCalibration
Calibration
• Perform auto-calibration by running a known ‘truth’ distance, such as a 400-meter track. For best results,
choose a truth distance between 400 and 1000 meters
• After running, adjust the measured distance to the ‘actual’ distance that you ran.
• Auto-calibration will automatically calculate and apply the required calibration factor for you.
• *** When calibrating, place the footpod on the same foot that you will typically use for all runs.
• *** Perform the run at your typical running pace. Note that the footpod will still give accurate data at other
speeds.