User Manual

Preliminary document not for redistribution. Company confidential. Product features and specifications may vary.
Copyright ©2006 Dynastream Innovations Inc. REV: 2006-08-18 Page 22
Footpod Calibration
The accuracy of the speed distance monitor (footpod) is 97% out of the box for a majority of users. Some users with
certain stride characteristics, such as pronation and supination, will not obtain 97% accuracy until a simple
calibration procedure is performed. Calibration can be done either manually or using the auto-calibration feature.
Manual Calibration
Manual CalibrationManual Calibration
Manual Calibration
Manual calibration involves the adjustment of a multiplication factor that is then applied to all future speed
and distance measurements.
The nominal value of the manual calibration factor is 1.000.
The nominal value can be adjusted upwards or downwards if the athlete has a known reference distance to
calculate the calibration factor appropriately.
The manual calibration factor will be automatically adjusted following an auto-calibration process, and can be
recorded for future reference (for example to track calibration values for multiple users).
Auto
AutoAuto
Auto-
--
-Calibration
CalibrationCalibration
Calibration
Perform auto-calibration by running a known ‘truth’ distance, such as a 400-meter track. For best results,
choose a truth distance between 400 and 1000 meters
After running, adjust the measured distance to the ‘actual’ distance that you ran.
Auto-calibration will automatically calculate and apply the required calibration factor for you.
*** When calibrating, place the footpod on the same foot that you will typically use for all runs.
*** Perform the run at your typical running pace. Note that the footpod will still give accurate data at other
speeds.