Instruction Manual

Rev. 5/09 - 2.5 - TFXP
Table 2.2 Transducer Mounting Modes - DTTN / DTTH
Transducer
Mounting Mode
Pipe Material Pipe Size Liquid Composition
W-Mount
Plastic (all types)
Carbon Steel
Stainless Steel
Copper
Ductile Iron
Cast Iron1
2-4 in (50-100 mm)
2-4 in (50-100 mm)
2-4 in (50-100 mm)
2-4 in (50-100 mm)
Not recommended
Not recommended
Low TSS; non-aerated
V-Mount
Plastic (all types)
Carbon Steel
Stainless Steel
Copper
Ductile Iron
Cast Iron
4-12 in (100-300 mm)
4-12 in (100-300 mm)
4-12 in (100-300 mm)
4-30 in (100-750 mm)
2-12 in (50-300 mm)
2-12 in (50-300 mm)
Low TSS; non-aerated
Z-Mount
Plastic (all types)
Carbon Steel
Stainless Steel
Copper
Ductile Iron
Cast Iron
> 30 in (> 750 mm)
> 12 in (> 300 mm)
> 12 in (> 300 mm)
> 30 in (> 750 mm)
> 12 in (> 300 mm)
> 12 in (> 300 mm)
Low TSS; non-aerated
TSS = Total Suspended Solids
Figure 2.1 Transducer Mounting Modes - DTTN / DTTH
PART 2 - TRANSDUCER & RTD INSTALLATION