User Manual

TFX Error Codes
Revised 2-22-20 02
Code Number Description Correction
Warnings
0001
Serial number not present Hardware serial number has become inoperative – system
performance will not be influenced.
0010
Signal Strength is below Signal
Strength Cutoff entry
Low signal strength is typically caused by one of the following:
Empty pipe
Improper programming/incorrect values
Improper transducer spacing
Non-homogeneous pipe wall
0011
Measured Speed of Sound the in the
liquid is greater than 10% different than
the value entered during meter setup
Verify that the correct liquid was selected in the BASIC menu.
Verify that pipe size parameters are correct.
0020
Heat Flow Units of measure have been
selected and an RTD module has not
been installed
Verify that RTD Module PN D020-1045-106 has been installed in
one of the I/O meter slots. Verify that OUTPUT1 or OUTPUT 2 has
been configured for RTD measurements.
Class C Errors
1001
System tables have changed Initiate a meter RESET by cycling power or by selecting SYSTEM
RESET in the SEC MENU.
1002
System configuration has changed Initiate a meter RESET by cycling power or by selecting SYSTEM
RESET in the SEC MENU.
Class B Errors
3001
Invalid hardw are configuration Upload correct ed file
3002
Invalid syste m conf igura tion Upload correct ed file
3003
Invalid strategy file Upload correct ed file
3004
Invalid calibration data Recalibrate the system
3005
Invalid speed of sound calibration data Upload new data
3006
Bad system tables Upload new table data
3007
Data Logger is off or not present If desired, insert data logger and configure within the Datalog
Operations Menu. If logger is not present, configure I/O port for no
logger.
3010
One or more channels are not
responding (Multi-channel meters only)
Display indicates which secondary units are not communicating
with Master meter. Verify wiring, configuration and address of
secondary instrument.
3011
All channels are not responding (Multi-
channel met ers only )
Verify wiring, configuration and address of secondary instruments.
Class A Errors
4001
Flash memory full Return unit to factory for evaluation