Manual
Rev. 01/11 -2.6- MFX
PART 2 - MAGPROBE INSTALLATION
Figure 2.3
Installation
Measurements
Installation Measurements: DMP2 through
DMP6 MagProbes
TO CALCULATE INSERTION DEPTH
Measure and record the following linear
distances:
E = PROBE LENGTH = _______
C = INSERTION FITTING to PIPE WALL
= _______
B = PIPE WALL THICKNESS
= _______
A = 0.125 x PIPE ID = _______
D = INSERTION DEPTH = _______
D = E - C - B - A