Instruction Manual

23
2. Press the key to display FL C-OFF on bottom display. Press
the ENTER key.
3. Press the
key to move one digit right and the key to
increment a flashing digit. Values are entered in flow rate
units.
4. Press the ENTER key.
5. Press the MENU key to exit.
OUTPUT -- Output Type
NONE - 4-20mA and Rate Pulse Outputs disabled
4-20MA - 4-20mA Analog Output
RATE - Rate Pulse Output (0-1,000 Hz)
Utilized to select either the 4-20mA or 1,000 Hz rate pulse
output.
1. To select either the mA or pulse output, press the MENU key.
2. Press the
key to display output on bottom display. Press
ENTER key.
3. Press the
key to select 4-20mA, pulse or none.
4. Press the ENTER key.
5. Press the MENU key to exit.
420TEST -- 4-20 mA Output Test
Allows a fixed current to be output from the 4-20 mA output. By
incrementing this value, the 4-20 mA output will transmit the
indicated current value.
The 4-20TEST prompt only appears if the output is
configured for 4-20mA output. The output will be spanned
between 0 flow and the value entered for Full Flow Rate.
Select the 420TEST feature. Use the
and keys to increment/
decrement mA output current. Verify that at mid-scale, 12 mA,
that the receiving unit is showing 50% of full flow.
IMPORTANT: The RATE PULSE and 4-20 mA outputs cannot
be operated simultaneously. Select the desired output type
from the OUTPUT menu.