Press Delta Time Zero (ns) zeros to remove the No Flow transit time offset. This is also referred to as Zeroing the meter.
Because every flowmeter installation is slightly different and sound waves can travel in slightly different ways through these
various installations, it is important to remove the zero offset at zero flow to maintain the meter’s accuracy. A provision is
made using this button to establish Zero flow and eliminate the offset.
Procedure:
1. The pipe must be full of liquid.
2. Flow must be absolutely zero. Securely close any valves and allow time for any settling to occur.
3. Press the Transit Time Zero button (Delta Time Zero) once.