INPUT / OUTPUT I/O GROUP
Display Adv Cal System
Site
Meter Log
I/O
Test OutSet Out Scale Out
Test Out
Set Out
Scale Out
Figure 83: I/O Group
Setup Outputs (Set Out)
Site Name0.00 GPM
3/28/2012
13:24
Site Adv Cal System
Display
I/O
Meter Log
Test Out
Scale Out
Set Out
Enable 4-20mA
Output
Enable 0-10V
Output
Enable 0-1kHz
Output
Figure 84: Data Outputs Setup
The I/O>Set Out controls which of the DXN outputs are active. The DXN output signals can be described as analog, or digital/
pulse. Analog signals change continuously over time, digital/pulse signals are present at discrete points and represent
information using a sequence of on/off pulses. For connection information see Figure 110.
The type of I/O available from the DXN is set using check boxes. The two analog outputs are 4-20 mA and 0…10V DC output.
The pulse output has a maximum frequency of 1000 Hz and can be configured as either a rate pulse or a totalizing pulse.
Whichever output is used all outputs are scaled the same way using the I/O>Scale Out. In other words 20 mA,
1000 Hz, and 0…10V will all represent the same Maximum flow rate set in the I/O>Scale Out.
Scale Outputs (Scale Out)
Site Name0.00 GPM
3/28/2012
13:24
Site Adv Cal System
Display
I/O
Meter Log
Test OutSet Out
Scale Out
Data To Output
Flow
100
Flow at Min Out
(GPM)
Flow at Max Out
(GPM)
-100
Figure 85: Scale Outputs Setup
DXN Portable Ultrasonic Measurement System
Page 66 February 2013