PART 4 - TROUBLE SHOOTING
FAULT light is ON; low SIGNAL
STRENGTH indication (continued)
If GND connection and pipe are at
different potentials, ground D901/M to
pipe potential.
If Variable Frequency Drives are being
utilized, verify that the D901/M obtains a
READ light when the pump turn OFF. If it
does, contact the Dynasonics factory.
Stability of flow readings are
unsatisfactory
Increase the DAMP constant from keypad.
Increase the system time constant by
turning R17 (DAMP) clockwise till
readings are satisfactory.
Move transducers to a location further
from piping tees, elbows, valves, filters,
etc.
Erroneous Reading
Transducers mounted incorrectly.
Another local ultrasonic instrument is
operating at about the same frequency
[ consult the Dynasonics factory].
Presence of large amounts of suspended
solids or aeration. Use CAL constant to
compensate.
Sources of radiated interference are
present. Apply appropriate shielding.
An electrically noisy power supply is
powering the D901/M. Power the meter
with a circuit that does not power motors,
ballasts or switching supplies.
Readings of flow at zero fluid velocity
Verify that residual leakage and flow is not
present. [I.e. leaking check valves]
Verify that SENSITIVITY is not adjusted
too high. With nominal flow running
through the pipe, adjust SENSITIVITY
control till the right-most bar meter light
just comes ON.