PART 4 - TROUBLE SHOOTING
CONDITION
POSSIBLE CAUSE
Unit does not turn “ON” when power is
applied
Check AC connections.
Test the fuse
Ensure the terminal block located in the
upper left corner of the main PCB is
secure
Verify that ribbon cables between PCBs
are connected.
OVERRANGE light is ON
Increase the value of the FULL SCALE
constant.
Verify that fluid velocity is not greater
than 20 FPS [6.08 MPS]
FAULT light is ON; low SIGNAL
STRENGTH indication
Ensure that the transducers are properly
mounted to the pipe.
Verify that transducer connections are
correct
Ensure that the pipe is full of moving
liquid.
Increase SENSITIVITY so that right-
most SIGNAL STRENGTH light just
comes ON.
On cleaner liquids, move the
transducers closer to a 90
o
pipe elbow.
On dirtier liquids, mount the transducers
as described in CASE 3 of Part 2 of this
manual.
If the pipe has a polyethylene liner,
move the transducers to another area.
The liner may contain an air void at this
location.
(continued)