Product Manual
Oximeter SerieSOximeter SerieS
PULSE OXIMETER INSTRUCTION MANUAL
42
Symbols
– minus; negative © copyright
% percent / per; divide; or
+ plus = equal to
< less than > greater than
≤ less than or equal to ≥ greater than or equal to
± plus or minus × multiply
APPENDIX B
Guidance and manufacturer’s declaration -
electromagnetic immunity
The Hand Held Pulse Oximeter is intended for use in
an electromagnetic environment specied below. The
customer or the user of the Pulse Oximeter should assure
that it is used in such an environment.
Immunity
Test
IEC
60601
Test
level
Compliance
Level
Electromagnetic
environment
guidance
Electro-
static
Discharge
(ESD)
IEC61004-2
6kV
contact
8kV air
6kV
contact
8kV air
Floors should be
wood, concrete
or ceramic tile. If
oor are convert-
ed with Synthetic
material, the
relative humidity
should be at
least 30%